Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Published June 10, 2026 · Recall initiated: December 7, 2023
Summary
FDA records medical device Class II recall Z-2289-2026, published June 10, 2026 and initiated December 7, 2023. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 156 units; Lot: UDI (01)00884838085367(21)547, (01)00884838099272(21)120, (01)00884838099272(21)122, (01)00884838099272(21)123, (01)00884838099272(21)124, (01)00884838099272(21)125, (01)00884838099272(21)126, (01)00884838099272(21)127, (01)00884838099272(21)153, (01)00884838099272(21)67, (01)00884838099272(21)209, (01)00884838099272(21)271, (01)00884838099272(21)675, (01)00884838099272(21)112, (01)00884838085367(21)467, (01)00884838085367(21)335, (01)00884838099272(21)113, (01)00884838099272(21)121, (01)00884838099272(21)139, (01)00884838099272(21)158, (01)00884838085367(21)455, (01)00884838085367(21)494, (01)00884838099272(21)239, (01)00884838099272(21)105, (01)00884838099272(21)678, (01)00884838099272(21)362, (01)00884838099272(21)114, (01)00884838085367(21)564, (01)00884838099272(21)403, (01)00884838085367(21)418, (01)00884838085367(21)440, (01)00884838099272(21)108, (01)00884838099272(21)545, (01)00884838099272(21)197, (01)00884838099272(21)380, (01)00884838099272(21)281, (01)00884838099272(21)495, (01)00884838099272(21)319, (01)00884838099272(21)339, (01)00884838099272(21)182, (01)00884838099272(21)183, (01)00884838085367(21)64, (01)00884838099272(21)325, (01)00884838099272(21)100, (01)00884838099272(21)272, (01)00884838099272(21)680, (01)00884838085367(21)42, (01)00884838099272(21)255, (01)00884838099272(21)341, (01)00884838099272(21)687, (01)00884838099272(21)350, (01)00884838099272(21)280, (01)00884838085367(21)323, (01)00884838099272(21)270, (01)00884838099272(21)254, (01)00884838099272(21)107, (01)00884838099272(21)144, (01)00884838085367(21)109, (01)00884838099272(21)294, (01)00884838099272(21)559, (01)00884838099272(21)98, (01)00884838099272(21)343, (01)00884838085367(21)443, (01)00884838099272(21)301, (01)00884838099272(21)117, (01)00884838085367(21)322, (01)00884838085367(21)373, (01)00884838085367(21)442, (01)00884838099272(21)244, (01)00884838099272(21)332, (01)00884838085367(21)308, (01)00884838099272(21)78, (01)00884838099272(21)374, (01)00884838099272(21)11, (01)00884838085367(21)354, (01)00884838099272(21)441, (01)00884838099272(21)129, (01)00884838099272(21)180, (01)00884838085367(21)279, (01)00884838099272(21)442, (01)00884838085367(21)400, (01)00884838099272(21)220, (01)00884838099272(21)189, (01)00884838099272(21)190, (01)00884838099272(21)446, (01)00884838085367(21)510, (01)00884838085367(21)111, (01)00884838085367(21)113, (01)00884838085367(21)446, (01)00884838099272(21)208, (01)00884838099272(21)554, (01)00884838085367(21)330, (01)00884838085367(21)288, (01)00884838099272(21)360, (01)00884838099272(21)337, (01)00884838099272(21)159, (01)00884838099272(21)326, (01)00884838099272(21)400, (01)00884838085367(21)329, (01)00884838099272(21)175, (01)00884838099272(21)427, (01)00884838099272(21)464, (01)00884838099272(21)468, (01)00884838085367(21)472, (01)00884838085367(21)470, (01)00884838085367(21)457, (01)00884838085367(21)520, (01)00884838085367(21)490, (01)00884838099272(21)143, (01)00884838099272(21)262, (01)00884838085367(21)103, (01)00884838099272(21)391, (01)00884838085367(21)150, (01)00884838085367(21)277, (01)00884838099272(21)173, (01)00884838085367(21)172, (01)00884838099272(21)147, (01)00884838099272(21)204, (01)00884838099272(21)467, (01)00884838085367(21)356, (01)00884838085367(21)130, (01)00884838099272(21)260, (01)00884838099272(21)381, (01)00884838099272(21)508, (01)00884838099272(21)419, (01)00884838099272(21)313, (01)00884838085367(21)469, (01)00884838099272(21)472, (01)00884838099272(21)318, (01)00884838099272(21)473, (01)00884838085367(21)431, (01)00884838099272(21)174, (01)00884838099272(21)383, (01)00884838099272(21)181, (01)00884838099272(21)463, (01)00884838099272(21)205, (01)00884838099272(21)59, (01)00884838099272(21)361, (01)00884838085367(21)114, (01)00884838099272(21)555, (01)00884838099272(21)379, (01)00884838085367(21)511, (01)00884838085367(21)477, (01)00884838099272(21)266, (01)00884838099272(21)398, (01)00884838099272(21)593, (01)00884838099272(21)75, (01)00884838099272(21)19, (01)00884. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
What to do
Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- US Nationwide distribution.
- Lot
- UDI (01)00884838085367(21)547, (01)00884838099272(21)120, (01)00884838099272(21)122, (01)00884838099272(21)123, (01)00884838099272(21)124, (01)00884838099272(21)125, (01)00884838099272(21)126, (01)00884838099272(21)127, (01)00884838099272(21)153, (01)00884838099272(21)67, (01)00884838099272(21)209, (01)00884838099272(21)271, (01)00884838099272(21)675, (01)00884838099272(21)112, (01)00884838085367(21)467, (01)00884838085367(21)335, (01)00884838099272(21)113, (01)00884838099272(21)121, (01)00884838099272(21)139, (01)00884838099272(21)158, (01)00884838085367(21)455, (01)00884838085367(21)494, (01)00884838099272(21)239, (01)00884838099272(21)105, (01)00884838099272(21)678, (01)00884838099272(21)362, (01)00884838099272(21)114, (01)00884838085367(21)564, (01)00884838099272(21)403, (01)00884838085367(21)418, (01)00884838085367(21)440, (01)00884838099272(21)108, (01)00884838099272(21)545, (01)00884838099272(21)197, (01)00884838099272(21)380, (01)00884838099272(21)281, (01)00884838099272(21)495, (01)00884838099272(21)319, (01)00884838099272(21)339, (01)00884838099272(21)182, (01)00884838099272(21)183, (01)00884838085367(21)64, (01)00884838099272(21)325, (01)00884838099272(21)100, (01)00884838099272(21)272, (01)00884838099272(21)680, (01)00884838085367(21)42, (01)00884838099272(21)255, (01)00884838099272(21)341, (01)00884838099272(21)687, (01)00884838099272(21)350, (01)00884838099272(21)280, (01)00884838085367(21)323, (01)00884838099272(21)270, (01)00884838099272(21)254, (01)00884838099272(21)107, (01)00884838099272(21)144, (01)00884838085367(21)109, (01)00884838099272(21)294, (01)00884838099272(21)559, (01)00884838099272(21)98, (01)00884838099272(21)343, (01)00884838085367(21)443, (01)00884838099272(21)301, (01)00884838099272(21)117, (01)00884838085367(21)322, (01)00884838085367(21)373, (01)00884838085367(21)442, (01)00884838099272(21)244, (01)00884838099272(21)332, (01)00884838085367(21)308, (01)00884838099272(21)78, (01)00884838099272(21)374, (01)00884838099272(21)11, (01)00884838085367(21)354, (01)00884838099272(21)441, (01)00884838099272(21)129, (01)00884838099272(21)180, (01)00884838085367(21)279, (01)00884838099272(21)442, (01)00884838085367(21)400, (01)00884838099272(21)220, (01)00884838099272(21)189, (01)00884838099272(21)190, (01)00884838099272(21)446, (01)00884838085367(21)510, (01)00884838085367(21)111, (01)00884838085367(21)113, (01)00884838085367(21)446, (01)00884838099272(21)208, (01)00884838099272(21)554, (01)00884838085367(21)330, (01)00884838085367(21)288, (01)00884838099272(21)360, (01)00884838099272(21)337, (01)00884838099272(21)159, (01)00884838099272(21)326, (01)00884838099272(21)400, (01)00884838085367(21)329, (01)00884838099272(21)175, (01)00884838099272(21)427, (01)00884838099272(21)464, (01)00884838099272(21)468, (01)00884838085367(21)472, (01)00884838085367(21)470, (01)00884838085367(21)457, (01)00884838085367(21)520, (01)00884838085367(21)490, (01)00884838099272(21)143, (01)00884838099272(21)262, (01)00884838085367(21)103, (01)00884838099272(21)391, (01)00884838085367(21)150, (01)00884838085367(21)277, (01)00884838099272(21)173, (01)00884838085367(21)172, (01)00884838099272(21)147, (01)00884838099272(21)204, (01)00884838099272(21)467, (01)00884838085367(21)356, (01)00884838085367(21)130, (01)00884838099272(21)260, (01)00884838099272(21)381, (01)00884838099272(21)508, (01)00884838099272(21)419, (01)00884838099272(21)313, (01)00884838085367(21)469, (01)00884838099272(21)472, (01)00884838099272(21)318, (01)00884838099272(21)473, (01)00884838085367(21)431, (01)00884838099272(21)174, (01)00884838099272(21)383, (01)00884838099272(21)181, (01)00884838099272(21)463, (01)00884838099272(21)205, (01)00884838099272(21)59, (01)00884838099272(21)361, (01)00884838085367(21)114, (01)00884838099272(21)555, (01)00884838099272(21)379, (01)00884838085367(21)511, (01)00884838085367(21)477, (01)00884838099272(21)266, (01)00884838099272(21)398, (01)00884838099272(21)593, (01)00884838099272(21)75, (01)00884838099272(21)19, (01)00884
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAPhilips Azurion 5 M12. Model Number: 722227.November 4, 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.November 4, 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.November 4, 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.November 4, 2026
- FDAPhilips Azurion 7B12/12, Model Numbers: 722067, 722225 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5June 10, 2026
- FDAPhilips Azurion 7M12, Model Numbers: 722078, 722223 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 andJune 10, 2026
Questions about this recall
Was Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key recalled?
Yes. FDA published recall Z-2289-2026 on June 10, 2026. Reported hazard: During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
What should I do if I have this product?
Letter