FDAFDA · Food and drugs·Recall number Z-2310-2015
ANSPACH - 5MM Diamond Ball, 2MM Shaft Exposure; REF S-5D-2-G1 Cutting shaping bone including spine and cranium.
Published August 5, 2015 · Recall initiated: April 13, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- The Anspach Effort, Inc.
- Distributor
- Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
- Lot
- REF S-5D-2-G1 ; ALL lot numbers.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States