Willscher Vas Reapproximator Plate, Surgical Instrument, Supplied sterilized by ethylene oxide gas in peel-open packages. Used to stabilize the vas ends during microscopic vasovasostomy.

Published June 14, 2017 · Recall initiated: May 1, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

COOK Medical is initiating a voluntary recall of multiple products because the reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Cook Inc.
Distributor
Worldwide Distribution - US (nationwide) and countries of: Australia, Austria, Belgium, Bahrain, Canada, China, Costa Rica, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Lithuania, Latvia, Malaysia, Netherlands, Romania, Russia, Saudi Arabia, Singapore, Spain, Switzerland, United Arab Emirates, and United Kingdom.
Lot
CATALOG NUMBER: 092002 and 096511, GPN: G15612 and G15573, The device was obsoleted on 04/30/2008
Risk classification
Class II
Area (FDA)
device
Country
United States