Plum Duo Precision IV Pump, 40002-0403
Published June 17, 2026 · Recall initiated: April 29, 2026
Summary
FDA records medical device Class II recall Z-2313-2026, published June 17, 2026 and initiated April 29, 2026. Recalling firm: ICU Medical, Inc.; Status: Ongoing; Reported quantity: 13,613; Lot: UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
- Lot
- UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAIV Sets with drip chambersSeptember 23, 2026
- FDAOncology sets with drip chambersSeptember 23, 2026
- FDAIV Non-Dedicated Gravity Administration Sets with drip chambersSeptember 23, 2026
- FDAPlum Duo Infusion Pump, 40002-0401June 17, 2026
- FDAAllia IGS 3 Pulse angiographic X-ray systemJune 17, 2026
Questions about this recall
Was Plum Duo Precision IV Pump, 40002-0403 recalled?
Yes. FDA published recall Z-2313-2026 on June 17, 2026. Reported hazard: Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
What should I do if I have this product?
Voluntary: Firm initiated · Letter