Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging.
Published September 12, 2012 · Recall initiated: February 21, 2012
Summary
FDA records medical device Class II recall Z-2315-2012, published September 12, 2012 and initiated February 21, 2012. Recalling firm: Toshiba American Medical Systems Inc; Status: Terminated; Reported quantity: 3 units; Lot: MRT-1503/P5-Vantage MRI System. Serial Codes: P5A0632002, P5B06Y2005, P5B0712006, P3F11Z2143. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Toshiba American Medical Systems Inc
- Distributor
- US (nationwide) distribution including the states of Ohio and Texas.
- Lot
- MRT-1503/P5-Vantage MRI System. Serial Codes: P5A0632002, P5B06Y2005, P5B0712006, P3F11Z2143.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINFX-8000C Fluoroscopic X-Ray SystemsMarch 14, 2018
- FDAAPLIO 500 TUS-A500 diagnostic ultrasound systemJuly 19, 2017
- FDAXARIO 200 TUS-X200 diagnostic ultrasound systemJuly 19, 2017
- FDAToshiba Kalare DREX-KL80 X-ray generatorMarch 22, 2017
- FDATxRx Knee 15 Coil 3T. For use in conjunction with magnetic resonance scanners to produce diagnostic images ofSeptember 12, 2012
- FDABD GeneOhm MRSA ACP Assay Catalog #441637, box 48 tests labeled in part***GeneOhm Sciences Canada, Inc., 2555September 12, 2012
Questions about this recall
Was Toshiba Ultra-Short Magnetic Resonance Imaging System, EXCELART Vantage. MRT-1503/P5-Vantage MRI System. MR imaging. recalled?
Yes. FDA published recall Z-2315-2012 on September 12, 2012. Reported hazard: The firm recalled due to a potentail problem with the position of the hold in the lug teminals on the power cables connected to the terminal blocks on the gradient coil.
What should I do if I have this product?
Voluntary: Firm initiated · Letter