MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display
Published July 17, 2024 · Recall initiated: April 24, 2024
Summary
FDA records medical device Class II recall Z-2332-2024, published July 17, 2024 and initiated April 24, 2024. Recalling firm: Barco N.V.; Status: Completed; Reported quantity: 88 units; Lot: serial numbers: 2530986299, 2530986300, 2530986301, 2530986302, 2530986303, 2530986304, 2530986305, 2530986306, 2530986307, 2530986308, 2530986309, 2530986310, 2530986311, 2530986312, 2530986313, 2530986314, 2530986315, 2530986316, 2530986317, 2530986318, 2530986319, 2530986320, 2530986321, 2530986322, 2530986323, 2530986324, 2530986325, 2530986326, 2530986327, 2531041594, 2531041591, 2531041578, 2531041585, 2531041595, 2531041596, 2531041586, 2531041588, 2531041587, 2531041577, 2531041584, 2531041581, 2531041612, 2531041608, 2531041600, 2531041601, 2531041604, 2531041599, 2531041602, 2531041605, 2531041606, 2531041607, 2531041609, 2531041610, 2531041580, 2531041583, 2531041598, 2531041593, 2531041590, 2531041614, 2531041582, 2531041613, 2531041589, 2531041616, 2531041579, 2531041615, 2531041603, 2531041597, 2531041592, 2530985205, 2530986280, 2530986281, 2530986282, 2530986283, 2530986284, 2530986285, 2530986286, 2530986287, 2530986288, 2530986289, 2530986290, 2530986291, 2530986292, 2530986293, 2530986294, 2530986295, 2530986296, 2530986297, 2530986298. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Completed — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Barco N.V.
- Distributor
- US: FL
- Lot
- serial numbers: 2530986299, 2530986300, 2530986301, 2530986302, 2530986303, 2530986304, 2530986305, 2530986306, 2530986307, 2530986308, 2530986309, 2530986310, 2530986311, 2530986312, 2530986313, 2530986314, 2530986315, 2530986316, 2530986317, 2530986318, 2530986319, 2530986320, 2530986321, 2530986322, 2530986323, 2530986324, 2530986325, 2530986326, 2530986327, 2531041594, 2531041591, 2531041578, 2531041585, 2531041595, 2531041596, 2531041586, 2531041588, 2531041587, 2531041577, 2531041584, 2531041581, 2531041612, 2531041608, 2531041600, 2531041601, 2531041604, 2531041599, 2531041602, 2531041605, 2531041606, 2531041607, 2531041609, 2531041610, 2531041580, 2531041583, 2531041598, 2531041593, 2531041590, 2531041614, 2531041582, 2531041613, 2531041589, 2531041616, 2531041579, 2531041615, 2531041603, 2531041597, 2531041592, 2530985205, 2530986280, 2530986281, 2530986282, 2530986283, 2530986284, 2530986285, 2530986286, 2530986287, 2530986288, 2530986289, 2530986290, 2530986291, 2530986292, 2530986293, 2530986294, 2530986295, 2530986296, 2530986297, 2530986298
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Belgium
Related recalls
- FDAAURORA Surgiscope System. Sterile, single use device that contains a Sheath, Obturator, and Imager.July 17, 2024
- FDASynergetics I Pack Injection Kit- Tray for use with Ophthalmic injections Model/Catalog Number: 18060July 17, 2024
- FDAPresource kits and trays, labeled as: 1) Neurological Tray, Catalog Number SAN29NICCA; 2) Neurological Tray,July 17, 2024
- FDASynergetics I-Pack Injection Kit with Drape -For use with Ophthalmic injections Model/Catalog Number: 18061July 17, 2024
- FDAPresource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC3; 2)July 17, 2024
- FDABarco MDSC-8527 NXF, Part Number K9352421February 14, 2024
Questions about this recall
Was MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display recalled?
Yes. FDA published recall Z-2332-2024 on July 17, 2024. Reported hazard: We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail