STERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model.
Published September 19, 2012 · Recall initiated: June 19, 2012
Summary
FDA records medical device Class III recall Z-2337-2012, published September 19, 2012 and initiated June 19, 2012. Recalling firm: Advanced Sterilization Products; Status: Terminated; Reported quantity: 7,495 units; Lot: Product Code: 10113. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Advanced Sterilization Products
- Distributor
- Nationwide Distribution.
- Lot
- Product Code: 10113
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASTERRAD NX Sterilization System, Product Code: 10033, 10033-002February 22, 2017
- FDASTERRAD 100S, Product Code 10113. The STERRAD Cassettes are used on STERRAD Sterilization System which is aDecember 31, 2014
- FDASTERRAD NX, Product Code 10133 The STERRAD Cassettes are used on STERRAD Sterilization System which is a lowDecember 31, 2014
- FDASTERRAD 200, Product Code 10118 The STERRAD Cassettes are used on STERRAD Sterilization System which is a lowDecember 31, 2014
- FDABLOM-SINGER indwelling TEP Occluder, catalog IN2004-TO The Blom-Singer Indwelling (TEP(tracheoesophagealSeptember 19, 2012
- FDASTERRAD NX Cassette, Batch # 12A078. The STERRAD 100S and STERRAD NX sterilizations systems areSeptember 19, 2012
Questions about this recall
Was STERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model. recalled?
Yes. FDA published recall Z-2337-2012 on September 19, 2012. Reported hazard: Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter