FDAFDA · Food and drugs·Recall number Z-2351-2024
BD Multitest 6-Color TBNK CE-IVD, REF: 644611; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes CE-IVD, REF: 337166; BD Multitest 6-Color TBNK, IVD, REF: 662967; BD Multitest 6-Color TBNK Kit with BD Trucount Tubes IVD, REF: 662995. BD Multitest" 6-Color TBNK reagent with optional BD Trucount" Tubes is a six-color direct immunofluorescence reagent.
Published July 24, 2024 · Recall initiated: May 8, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Multitest with compromised raw material used to assess immune system status may result in flow cytometry plot with unusual staining pattern, nonspecific aggregates, high background, double positive event increase on CD3+CD4+vsCD3+CD8+ and/or CD3-CD16+CD56+vsCD3-CD19+, contaminated antibodies, may lead to erroneous or delayed results, additional patient sample, might impact healthcare decisions
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Becton, Dickinson and Company, BD Biosciences
- Distributor
- US: MN, NJ, NY, MD, UT, FL, NC, MA, KY, NE, NV, CA, LA, DC, OH, MI, PA, WA, CT, AL, DE, TN, IN, OR, IL, WI, NH, TX, GA, RI, SD, ID, NM, AR. OUS: CA, SG, NZ, AU, KR, IN, PE, GT, AR, MX, TW, CN, NI, CR, JP, BR, TH, BE, VN
- Lot
- REF/UDI-DI/Lot-Expiration: 644611/382906446114/20010-31-Aug-24, 49510-30-Sep-24, 32674-30-Sep-24; 337166/382903371662/17569-31-Jul-24, 22709-31-Aug-24, 49517-30-Sep-24; 662967/382906629678/17572-31-Jul-24, 17582-31-Jul-24, 26797-31-Aug-24, 69761-30-Sep-24, 49506-30-Sep-24; 662995/382906629951/49508-30-Sep-24
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States