SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02

Published June 17, 2026 · Recall initiated: May 4, 2026

Summary

FDA records medical device Class II recall Z-2353-2026, published June 17, 2026 and initiated May 4, 2026. Recalling firm: Micromed S.p.A.; Status: Ongoing; Reported quantity: 207 units; Lot: UDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Micromed S.p.A.
Distributor
Worldwide - US Nationwide distribution in the states of FL, PA.
Lot
UDI-DI: 8033928120897; Lot Code: Model No SD LTM 64 PLUS; UDI-DI 8033928120897; Serial Numbers BAA-0051/01-19, BAA-0069/02-19, BAA-0070/02-19, BAA-0071/02-19, BAA-0082/02-19, BAA-0083/02-19, BAA-0084/02-19, BAA-0085/02-19, BAA-0086/02-19, BAA-0087/02-19, BAA-0088/02-19, BAA-0089/02-19, BAA-0090/02-19, BAA-0091/02-19, BAA-0092/02-19, BAA-0093/02-19, BAA-0095/02-19, BAA-0096/02-19, BAA-0098/02-19, BAA-0104/02-19, BAA-0105/02-19, BAA-0106/02-19, BAA-0107/02-19, BAA-0108/02-19, BAA-0109/02-19, BAA-0110/02-19, BAA-0111/02-19, BAA-0112/02-19, BAA-0113/02-19, BAA-0115/02-19, BAA-0116/02-19, BAA-0117/02-19, BAA-0118/02-19, BAA-0119/02-19, BAA-0121/02-19, BAA-0122/02-19, BAA-0142/03-19, BAA-0155/03-20, BAA-0156/03-20, BAA-0191/04-20, BAA-0192/04-20, BAA-0193/04-20, BAA-0194/04-20, BAA-0286/04-21, BAA-0287/04-21, BAA-0288/04-21, BAA-0289/04-21, BAA-0329/04-22
Risk classification
Class II
Area (FDA)
device
Country
Italy

Related recalls

Questions about this recall

Was SD LTM 64 PLUS, EEG Amplifier/recorder, Model/Catalog Number: SD LTM 64 PLUS, Software Version: firmware 2021.02, 2022.01 or 2022.02 recalled?

Yes. FDA published recall Z-2353-2026 on June 17, 2026. Reported hazard: Natus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter