Instrumentation Laboratory ACL TOP 700 PN 0000280010 ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters

Published September 26, 2012 · Recall initiated: August 3, 2012

Summary

FDA records medical device Class II recall Z-2388-2012, published September 26, 2012 and initiated August 3, 2012. Recalling firm: Instrumentation Laboratory Co.; Status: Terminated; Reported quantity: 304 units; Lot: Serial Number range: 09080100 through12060403 Distributed between: Aug 2009 to Jun 2012. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Instrumentation Laboratory Co.
Distributor
Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BRAZIL, BRUNEI DARUSSALAM, BULGARIA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF LUXEMBOURG, MACAU, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PANAMA, PERU, POLAND, PORTUGAL, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SLOVAKIA (Slovak Republic), SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, TURKEY, UNITED KINGDOM, URUGUAY and VIETNAM.
Lot
Serial Number range: 09080100 through12060403 Distributed between: Aug 2009 to Jun 2012
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Instrumentation Laboratory ACL TOP 700 PN 0000280010 ACL TOP is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters recalled?

Yes. FDA published recall Z-2388-2012 on September 26, 2012. Reported hazard: Low Abnormal Fibrinogen Sample Reports as Normal and High Fibrinogen Sample Reports as Low

What should I do if I have this product?

Voluntary: Firm initiated · Letter