FDAFDA · Food and drugs·Recall number Z-2394-2026

Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPK; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKB; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKC; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPK; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKB; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKC; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPK; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPKB; PI MIDLINE 1L: 4FR X 20CM/CDC-32041-MPKC; PI MIDLINE 2-L: 5FR X 20CM/CDC-32052-MPKB; PI PICC NaviCurve: 1L 4FR x 55CM w BP/DLX-35541-CURVB; PI PICC: 1L 4FR x 55CM w Biopatch/DLX-35541-HPKB; PI PICC: 1L 4FR x 55CM w CHG/DLX-35541-HPKC; PI PICC: 2L 5FR x 55CM w Biopatch/DLX-35552-HPKB; PI PICC: 2L 5FR X 55CM w CHG/DLX-35552-HPKC; PI PICC: 3L 6FR x 55CM w Biopatch/DLX-35563-HPKB; AGBA PICC/DELTA KIT: 1L 4.5 FR X 40 CM/ASK-44041-NS; AGBA PICC/DELTA FG: 2L 5.5 FR X 40 CM/ASK-44052-NS; AGBA PICC/DELTA KIT: 1L 4.5 FR X 50 CM/ASK-45041-NS; PI PICC KIT: 2L 5.5 FR X 50 CM/ASK-45052-TG; PI AGBA PICC KIT: 3L 6 FR X 50 CM W/VPS/ASK-45063-NS; PICC KIT: 1L 4.5 FR X 55 CM CG+ VPS/ASK-45541-RH; AGBA PICC: 1L 4.5FR x 55CM w Teg CHG/DLX-45541-HPKC; AGBA PICC: 2L 5.5FR x 55CM w Teg CHG/DLX-45552-HPKC; AGBA PICC NaviCurve: 3L 6FR x 55CM TCG/DLX-45563-CURVC; AGBA PICC: 3L 6FR x 55CM w Biopatch/DLX-45563-HPKB; AGBA PICC: 3L 6FR x 55CM w Teg CHG/DLX-45563-HPKC; PI PICC KIT: 1-L 4 FR X 55 CM TIPTRACKER/CDC-35541-TTS; PI PICC NaviCurve: 1L 4FR x 55CM w CHG/DLX-35541-CURVC; PI PICC G4 Stylet: 1L 4FR x 55CM w BP/DLX-35541-VPSB; PI PICC NaviCurve: 2L 5FR x 55CM w CHG/DLX-35552-CURVC; PI PICC G4 Stylet: 2L 5FR x 55CM w BP/DLX-35552-VPSB; PI PICC G4 Stylet: 2L 5FR x 55CM w CHG/DLX-35552-VPSC; PI PICC NaviCurve: 3L 6FR x 55CM w CHG/DLX-35563-CURVC; PI PICC G4 Stylet: 3L 6FR x 55CM w CHG/DLX-35563-VPSC; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 40 CM/CDC-44041-VPS2; PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UWH1; PI AGBA PICC KIT: 2L 5.5FR X 55CM W/ VPS/ASK-45552-UWH1; AGBA PICC G4 Stylet: 1L 4.5FR x 55CM BP/DLX-45541-VPSB; ACCESS TRAY/ASK-04001-NWM; ACCESS TRAY/ASK-04001-VCUH3; PI PICC:1L 4FR X 40CM PRELOAD VPS STYLET/CDC-34041-VPS; PI PICC:2L 5FR X 50CM PRELOAD VPS STYLET/CDC-35052-VPS; ACCESS KIT/ASK-04001-BWK2; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU11; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PVD; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16CM/ASK-12703-LMDTG1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8IN/ASK-15703-ECMC4; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-15703-PNM; CVC KIT: 3 LUMEN 12 FR X 6IN/ASK-22123-UAB; 2-L CVC KIT: 4 FR X13 CM/CDC-24402-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-X1A; CVC KIT: 3L 7 FR X 16 CM/ASK-42703-CPR; CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-DMC2; CVC KIT: 2 LUMEN 8 FR X 8 IN (20 CM)/ASK-45802-PCMH1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-MM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHUMC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PSAU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-RHR1; PI CATHETER 2L: 8 FR X 16 CM AGB+/ASK-42802-DU1; PI CVC KIT: 2L 8 FR X 6 IN/ASK-42802-PHF4; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PHHN1; MULTILUMEN CVC KIT: 2L 8 FR X 6 IN/ASK-42802-POSU3; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/ASK-42802-PUCD1; MULTI-LUMEN CVC KIT:2-L 8FR X 6IN (16CM)/ASK-42802-PWHC1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UHC; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-MCY1; PI CVC KIT: 3L 7FR X 8IN (20 CM) AGB+/ASK-45703-BCH; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-EMC; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-MHMC1; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/ASK-45703-NWM; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCOR; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PIMH; MULTI LUMEN CVC KIT: 3L

Published June 24, 2026 · Recall initiated: April 21, 2026

Summary

FDA records medical device Class I recall Z-2394-2026, published June 24, 2026 and initiated April 21, 2026. Recalling firm: ARROW INTERNATIONAL, LLC; Status: Ongoing; Reported quantity: 177,014; Lot: REF(Material)/UDI-DI/Lot(Batch): CDC-02031-MK1A/10801902193275/33F25C0444; CDC-41541-MPK1A/10801902193275/33F24L1137; CDC-41552-MPK1A/10801902193275, 50801902130162/33F24L0382, 33F25A0315, 33F25B0630, 33F25K0421, 33S24B0571; ASK-41541-VFHM/10801902193275, 10801902218862/33F24L1080, 33F25B0475, 33F25C0351, 33F25D0404, 33F25E0556, 33F25E0784, 33F25F0186, 33F25J0218, 33F25K0077, 33F25K0100, 33F25M0182, 33F26A1549, 33F26B0823; CDC-41541-MPK/10801902193275, 10801902193275/33F25H0768, 33S23K0382; CDC-41541-MPKB/10801902193275, 10801902205411/33F24H0429, 33F24K0043, 33F24M0067, 33F24M0149, 33F25A0390, 33F25B0099, 33F25B0343, 33F25C0578, 33F25E0421, 33F25F0401, 33F25G0515, 33F25G0710, 33F25H0817, 33F25J1008, 33F25L0015, 33F25L0050, 33F25L0295, 33F25M0499, 33F26A0330, 33X25J0003; CDC-41541-MPKC/10801902205435, 10801902193275/33F24D0982, 33F24F0166, 33F24F0167, 33F24H0398, 33F24H0454, 33F24J0380, 33F24J0389, 33F24K0581, 33F24L0182, 33F24M0022, 33F24M0077, 33F25A0348, 33F25B0228, 33F25B0281, 33F25B0289, 33F25B0344, 33F25C0513, 33F25C0532, 33F25C0558, 33F25D0710, 33F25D0736, 33F25F0020, 33F25F0025, 33F25F0032, 33F25H0818, 33F25J0577, 33F25J0657, 33F25L0251, 33F25L0296, 33S25A0211, 33F26C0435; CDC-41552-MPK/10801902201505/33F24D0561, 33F26B0490, 33S24A0521; CDC-41552-MPKB/10801902193275, 10801902205428/33F24J0183, 33F24J0221, 33F24L0010, 33F24L0173, 33F24M0114, 33F25B0223, 33F25B0299, 33F25D0724, 33F25F0436, 33F25G0640, 33F25K0354, 33F25L0265, 33F26B1048; CDC-41552-MPKC/10801902205442, 10801902193275/33F24C0861, 33F24D0343, 33F24F0090, 33F24K0474, 33F25C0208, 33F25D0559, 33F25E0827, 33F25H0608, 33F25J0434, 33F25M0248, 33F26A1559, 33F26C0776; CDC-32041-MPK/10801902193275, 10801902201475/33F25C0701, 33F25H0613, 33F25J0433, 33F25L0115, 33F25M0250, 33F26B0436; CDC-32041-MPKB/10801902205398, 10801902193275/33F24D0854, 33F24F0897, 33F24L0952, 33F25B0502, 33F25D0512, 33F25E0825, 33F25G0327, 33F25J0327, 33F25M0210, 33F26B0434, 33S23E0870; CDC-32041-MPKC/10801902193275/33F25D0513, 33F25H0590, 33F26A1387; CDC-32052-MPKB/10801902205404, 10801902193275/33F24F0825, 33F24L0381, 33F25D0514, 33F25H0609; DLX-35541-CURVB/10801902193275, 10801902208757/33F25D0674, 33F25J0186, 33S23G0883, 33S23H0988; DLX-35541-HPKB/10801902193275/33F25D0129, 33F25E0708; DLX-35541-HPKC/10801902193275/33F25D0244, 33F25E0830; DLX-35552-HPKB/10801902193275/33F25D0246, 33F25D0518, 33F25F0149; DLX-35552-HPKC/10801902193275, 10801902216721/33F25D0247, 33F25D0519, 33F25H0460, 33S23K0380; DLX-35563-HPKB/10801902193275/33F25K0318; ASK-44041-NS/10801902193275, 10801902216462/33F24L0347, 33S23H0298; ASK-44052-NS/10801902193275, 10801902216455/33F25A0302, 33S23J0265; ASK-45041-NS/10801902216677/33S24A0529; ASK-45052-TG/10801902193275/33F24L0864, 33F25D0409, 33F25H0490, 33F26A1379; ASK-45063-NS/10801902193275/33F24L1002; ASK-45541-RH/10801902193275/33F25B0574, 33F25E0607, 33F25H0427, 33F25J0161, 33F26A1412; DLX-45541-HPKC/10801902208849, 10801902193275/33F24D0604, 33F25D0128, 33F25E0763, 33F25J0131, 33F25M0315, 33F26B0456; DLX-45552-HPKC/10801902208856, 10801902193275/33F24F0097, 33F25C0450, 33F25D0252, 33F25D0625, 33F25G0321, 33F25H0601, 33F26B0213; DLX-45563-CURVC/10801902193275, 10801902208740/33F25A0321, 33F25B0635, 33F25D0522, 33F25E0766, 33F25G0394, 33F26C0707; DLX-45563-HPKB/10801902193275/33F25D0384, 33F25F0226; DLX-45563-HPKC/10801902208863, 10801902193275/33F24D0184, 33F25B0698, 33F25E0661, 33F26A1545; CDC-35541-TTS/10801902193275/33F25B0468; DLX-35541-CURVC/10801902193275, 10801902216684/33F25D0574, 33F25E0759, 33F25G0216, 33F25H0591, 33F25M0200, 33F26B0192; DLX-35541-VPSB/10801902193275, 10801902208665/33F25C0212, 33F25D0378, 33F25E0831, 33F25F0101, 33F25G0364, 33F25J0128, 33F25J0256, 33F25K0193, 33F25M0164, 33F26A1131, 33F26B0180, 33F26C0699; DLX-35552-CURVC/10801902193275, 10801902216714/33F25C0373, 33F25D0619, 33F25F0222, 33S24C0282; DLX-35552-VPSB/10801902193275, 10801902208672/33F25B0536, 33F25D0248, 33F25D0379, 33F25D0575, 33F25H0471, 33F2. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ARROW INTERNATIONAL, LLC
Distributor
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Lot
REF(Material)/UDI-DI/Lot(Batch): CDC-02031-MK1A/10801902193275/33F25C0444; CDC-41541-MPK1A/10801902193275/33F24L1137; CDC-41552-MPK1A/10801902193275, 50801902130162/33F24L0382, 33F25A0315, 33F25B0630, 33F25K0421, 33S24B0571; ASK-41541-VFHM/10801902193275, 10801902218862/33F24L1080, 33F25B0475, 33F25C0351, 33F25D0404, 33F25E0556, 33F25E0784, 33F25F0186, 33F25J0218, 33F25K0077, 33F25K0100, 33F25M0182, 33F26A1549, 33F26B0823; CDC-41541-MPK/10801902193275, 10801902193275/33F25H0768, 33S23K0382; CDC-41541-MPKB/10801902193275, 10801902205411/33F24H0429, 33F24K0043, 33F24M0067, 33F24M0149, 33F25A0390, 33F25B0099, 33F25B0343, 33F25C0578, 33F25E0421, 33F25F0401, 33F25G0515, 33F25G0710, 33F25H0817, 33F25J1008, 33F25L0015, 33F25L0050, 33F25L0295, 33F25M0499, 33F26A0330, 33X25J0003; CDC-41541-MPKC/10801902205435, 10801902193275/33F24D0982, 33F24F0166, 33F24F0167, 33F24H0398, 33F24H0454, 33F24J0380, 33F24J0389, 33F24K0581, 33F24L0182, 33F24M0022, 33F24M0077, 33F25A0348, 33F25B0228, 33F25B0281, 33F25B0289, 33F25B0344, 33F25C0513, 33F25C0532, 33F25C0558, 33F25D0710, 33F25D0736, 33F25F0020, 33F25F0025, 33F25F0032, 33F25H0818, 33F25J0577, 33F25J0657, 33F25L0251, 33F25L0296, 33S25A0211, 33F26C0435; CDC-41552-MPK/10801902201505/33F24D0561, 33F26B0490, 33S24A0521; CDC-41552-MPKB/10801902193275, 10801902205428/33F24J0183, 33F24J0221, 33F24L0010, 33F24L0173, 33F24M0114, 33F25B0223, 33F25B0299, 33F25D0724, 33F25F0436, 33F25G0640, 33F25K0354, 33F25L0265, 33F26B1048; CDC-41552-MPKC/10801902205442, 10801902193275/33F24C0861, 33F24D0343, 33F24F0090, 33F24K0474, 33F25C0208, 33F25D0559, 33F25E0827, 33F25H0608, 33F25J0434, 33F25M0248, 33F26A1559, 33F26C0776; CDC-32041-MPK/10801902193275, 10801902201475/33F25C0701, 33F25H0613, 33F25J0433, 33F25L0115, 33F25M0250, 33F26B0436; CDC-32041-MPKB/10801902205398, 10801902193275/33F24D0854, 33F24F0897, 33F24L0952, 33F25B0502, 33F25D0512, 33F25E0825, 33F25G0327, 33F25J0327, 33F25M0210, 33F26B0434, 33S23E0870; CDC-32041-MPKC/10801902193275/33F25D0513, 33F25H0590, 33F26A1387; CDC-32052-MPKB/10801902205404, 10801902193275/33F24F0825, 33F24L0381, 33F25D0514, 33F25H0609; DLX-35541-CURVB/10801902193275, 10801902208757/33F25D0674, 33F25J0186, 33S23G0883, 33S23H0988; DLX-35541-HPKB/10801902193275/33F25D0129, 33F25E0708; DLX-35541-HPKC/10801902193275/33F25D0244, 33F25E0830; DLX-35552-HPKB/10801902193275/33F25D0246, 33F25D0518, 33F25F0149; DLX-35552-HPKC/10801902193275, 10801902216721/33F25D0247, 33F25D0519, 33F25H0460, 33S23K0380; DLX-35563-HPKB/10801902193275/33F25K0318; ASK-44041-NS/10801902193275, 10801902216462/33F24L0347, 33S23H0298; ASK-44052-NS/10801902193275, 10801902216455/33F25A0302, 33S23J0265; ASK-45041-NS/10801902216677/33S24A0529; ASK-45052-TG/10801902193275/33F24L0864, 33F25D0409, 33F25H0490, 33F26A1379; ASK-45063-NS/10801902193275/33F24L1002; ASK-45541-RH/10801902193275/33F25B0574, 33F25E0607, 33F25H0427, 33F25J0161, 33F26A1412; DLX-45541-HPKC/10801902208849, 10801902193275/33F24D0604, 33F25D0128, 33F25E0763, 33F25J0131, 33F25M0315, 33F26B0456; DLX-45552-HPKC/10801902208856, 10801902193275/33F24F0097, 33F25C0450, 33F25D0252, 33F25D0625, 33F25G0321, 33F25H0601, 33F26B0213; DLX-45563-CURVC/10801902193275, 10801902208740/33F25A0321, 33F25B0635, 33F25D0522, 33F25E0766, 33F25G0394, 33F26C0707; DLX-45563-HPKB/10801902193275/33F25D0384, 33F25F0226; DLX-45563-HPKC/10801902208863, 10801902193275/33F24D0184, 33F25B0698, 33F25E0661, 33F26A1545; CDC-35541-TTS/10801902193275/33F25B0468; DLX-35541-CURVC/10801902193275, 10801902216684/33F25D0574, 33F25E0759, 33F25G0216, 33F25H0591, 33F25M0200, 33F26B0192; DLX-35541-VPSB/10801902193275, 10801902208665/33F25C0212, 33F25D0378, 33F25E0831, 33F25F0101, 33F25G0364, 33F25J0128, 33F25J0256, 33F25K0193, 33F25M0164, 33F26A1131, 33F26B0180, 33F26C0699; DLX-35552-CURVC/10801902193275, 10801902216714/33F25C0373, 33F25D0619, 33F25F0222, 33S24C0282; DLX-35552-VPSB/10801902193275, 10801902208672/33F25B0536, 33F25D0248, 33F25D0379, 33F25D0575, 33F25H0471, 33F2
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPK; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKB; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKC; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPK; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKB; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKC; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPK; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPKB; PI MIDLINE 1L: 4FR X 20CM/CDC-32041-MPKC; PI MIDLINE 2-L: 5FR X 20CM/CDC-32052-MPKB; PI PICC NaviCurve: 1L 4FR x 55CM w BP/DLX-35541-CURVB; PI PICC: 1L 4FR x 55CM w Biopatch/DLX-35541-HPKB; PI PICC: 1L 4FR x 55CM w CHG/DLX-35541-HPKC; PI PICC: 2L 5FR x 55CM w Biopatch/DLX-35552-HPKB; PI PICC: 2L 5FR X 55CM w CHG/DLX-35552-HPKC; PI PICC: 3L 6FR x 55CM w Biopatch/DLX-35563-HPKB; AGBA PICC/DELTA KIT: 1L 4.5 FR X 40 CM/ASK-44041-NS; AGBA PICC/DELTA FG: 2L 5.5 FR X 40 CM/ASK-44052-NS; AGBA PICC/DELTA KIT: 1L 4.5 FR X 50 CM/ASK-45041-NS; PI PICC KIT: 2L 5.5 FR X 50 CM/ASK-45052-TG; PI AGBA PICC KIT: 3L 6 FR X 50 CM W/VPS/ASK-45063-NS; PICC KIT: 1L 4.5 FR X 55 CM CG+ VPS/ASK-45541-RH; AGBA PICC: 1L 4.5FR x 55CM w Teg CHG/DLX-45541-HPKC; AGBA PICC: 2L 5.5FR x 55CM w Teg CHG/DLX-45552-HPKC; AGBA PICC NaviCurve: 3L 6FR x 55CM TCG/DLX-45563-CURVC; AGBA PICC: 3L 6FR x 55CM w Biopatch/DLX-45563-HPKB; AGBA PICC: 3L 6FR x 55CM w Teg CHG/DLX-45563-HPKC; PI PICC KIT: 1-L 4 FR X 55 CM TIPTRACKER/CDC-35541-TTS; PI PICC NaviCurve: 1L 4FR x 55CM w CHG/DLX-35541-CURVC; PI PICC G4 Stylet: 1L 4FR x 55CM w BP/DLX-35541-VPSB; PI PICC NaviCurve: 2L 5FR x 55CM w CHG/DLX-35552-CURVC; PI PICC G4 Stylet: 2L 5FR x 55CM w BP/DLX-35552-VPSB; PI PICC G4 Stylet: 2L 5FR x 55CM w CHG/DLX-35552-VPSC; PI PICC NaviCurve: 3L 6FR x 55CM w CHG/DLX-35563-CURVC; PI PICC G4 Stylet: 3L 6FR x 55CM w CHG/DLX-35563-VPSC; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 40 CM/CDC-44041-VPS2; PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UWH1; PI AGBA PICC KIT: 2L 5.5FR X 55CM W/ VPS/ASK-45552-UWH1; AGBA PICC G4 Stylet: 1L 4.5FR x 55CM BP/DLX-45541-VPSB; ACCESS TRAY/ASK-04001-NWM; ACCESS TRAY/ASK-04001-VCUH3; PI PICC:1L 4FR X 40CM PRELOAD VPS STYLET/CDC-34041-VPS; PI PICC:2L 5FR X 50CM PRELOAD VPS STYLET/CDC-35052-VPS; ACCESS KIT/ASK-04001-BWK2; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU11; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PVD; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16CM/ASK-12703-LMDTG1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8IN/ASK-15703-ECMC4; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-15703-PNM; CVC KIT: 3 LUMEN 12 FR X 6IN/ASK-22123-UAB; 2-L CVC KIT: 4 FR X13 CM/CDC-24402-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-X1A; CVC KIT: 3L 7 FR X 16 CM/ASK-42703-CPR; CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-DMC2; CVC KIT: 2 LUMEN 8 FR X 8 IN (20 CM)/ASK-45802-PCMH1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-MM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHUMC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PSAU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-RHR1; PI CATHETER 2L: 8 FR X 16 CM AGB+/ASK-42802-DU1; PI CVC KIT: 2L 8 FR X 6 IN/ASK-42802-PHF4; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PHHN1; MULTILUMEN CVC KIT: 2L 8 FR X 6 IN/ASK-42802-POSU3; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/ASK-42802-PUCD1; MULTI-LUMEN CVC KIT:2-L 8FR X 6IN (16CM)/ASK-42802-PWHC1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UHC; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-MCY1; PI CVC KIT: 3L 7FR X 8IN (20 CM) AGB+/ASK-45703-BCH; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-EMC; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-MHMC1; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/ASK-45703-NWM; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCOR; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PIMH; MULTI LUMEN CVC KIT: 3L recalled?

Yes. FDA published recall Z-2394-2026 on June 24, 2026. Reported hazard: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

What should I do if I have this product?

Voluntary: Firm initiated · Letter