MOSAIQ Oncology Information System

Published June 17, 2026 · Recall initiated: May 21, 2026

Summary

FDA records medical device Class II recall Z-2406-2026, published June 17, 2026 and initiated May 21, 2026. Recalling firm: Elekta, Inc.; Status: Ongoing; Reported quantity: 12 programs; Lot: Software Version: MOSAIQ 3.2/UDI: 07340201500071. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Elekta, Inc.
Distributor
Worldwide - US Nationwide distribution in the states of WI, FL, LA, NY, NY, NJ, KS, FL, IA, NM, TX and the countries of Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lithuania, Malta, Mexico, Netherlands, Pakistan, Panama, Poland, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom.
Lot
Software Version: MOSAIQ 3.2/UDI: 07340201500071
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MOSAIQ Oncology Information System recalled?

Yes. FDA published recall Z-2406-2026 on June 17, 2026. Reported hazard: Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.

What should I do if I have this product?

Voluntary: Firm initiated · Letter