Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399

Published July 24, 2024 · Recall initiated: May 23, 2024

Summary

FDA records medical device Class II recall Z-2412-2024, published July 24, 2024 and initiated May 23, 2024. Recalling firm: Alcon Research LLC; Status: Ongoing; Reported quantity: 230; Lot: UDI-DI: 00380650003997. Software Version: 1.5. Serial Numbers: 004-0680001, 004-0680002, 004-0680003, 004-0680004, 004-0680007, 004-0680008, 004-0680017, 004-0680026, 004-0680036, 004-0680047, 004-0680048, 004-0680086, 004-0680090, 004-0680092, 004-0680096, 004-0680100, 004-0680107, 004-0680112, 004-0680120, ,004-0680127,004-0680131, 004-0680134, 004-0680140, 004-0680152, 004-0680153, 004-0680161, 004-0680163, 004-0680188, 004-0680190,004-0680203, 004-0680206, 004-0680266, 004-0680267, 004-0680271, 004-0680273, 004-0680274, 004-0680283, 004-0680288, 004-0680293, 004-0680294, 004-0680306, 004-0680311, 004-0680315, 004-0680319, 004-0680325, 004-0680334, 004-0680336, 004-0680337, 004-0680338, 004-0680339, 004-0680346, 004-0680348, 004-0680352, 004-0680362, 004-0680404, 004-0680417, 004-0680418, 005-7230003, 005-7230005, 005-7230007, 005-7230017, 005-7230019, 005-7230021, 005-7230022, 005-7230023, 005-7230031, 005-7230035, 005-7230057, 005-7230081, 005-7230085, 005-7230086, 005-7230102, 005-7230108, 005-7230109, 005-7230186, 007-4610018, 007-4610023, 007-4610030, 007-4610032, 007-4610033, 007-4610034, 007-4610037, 007-4610044, 007-4610074, 007-4610107, 007-4610112, 007-4610115, 007-4610122, 007-4610123, 007-4610124, 007-4610125, 007-4610127, 007-4610132, 007-4610138, 007-4610166, 007-4610167, 007-4610179, 007-4610181, 007-4610186, 007-4610217, 007-4610219, 007-4610220, 007-4610226, 007-4610229, 007-4610230, 007-4610262, 007-4610275, 007-4610283, 007-4610287, 007-4610288, 007-4610308, 007-4610309, 007-4610351, 007-4610355, 007-4610359, 007-4610360, 007-4610366, 007-4610394, 007-4610404, 007-4610405, 007-4610406, 007-4610409, 009-4570001, 009-4570004, 009-4570005, 009-4570007, 009-4570013, 009-4570014, 009-4570015, 009-4570016, 009-4570019, 009-4570022, 009-4570033, 009-4570035, 009-4570044, 009-4570047, 009-4570048, 009-4570049, 009-4570051,009-4570052, 009-4570053, , 009-4570054,009-4570057, 009-4570059, 009-4570060, 009-4570064, 009-4570068, 009-4570070, 009-4570072, 009-4570073, 009-4570096, 009-4570097, 009-4570099, 009-4570100, 009-4570114, 009-4570116, 009-4570125, 009-4570127, 009-4570129, 009-4570130, 009-4570136, 009-4570137, 009-4570144, 009-4570148, 009-4570149, 009-4570150, 009-4570153, 009-4570155, 009-4570156, 009-4570158, 009-4570160, 009-4570161, 009-4570167, 009-4570171, 009-4570173, 009-4570176,009-4570178, 009-4570179, 009-4570180, 009-4570183, 009-4570184, 009-4570200, 009-4570204, 009-4570207, 009-4570233, 009-4570249, 009-4570256, 009-4570259, 009-4570263, 009-4570270, 009-4570275, 009-4570278, 009-4570281, 009-4570289, 009-4570293, 009-4570294, 009-4570300, 009-4570308, 009-4570317, 009-4570320, 009-4570331, 009-4570342, 009-4570343, 009-4570350, 009-4570353, 009-4570357, 009-4570365, 009-4570368, 009-4570393, 009-4570397, 009-4570425, 009-4570431, 009-4570435, 009-4570440, 009-4570448, 009-4570449, 009-4570451, 009-4570455, 009-4570457, 009-4570459, 009-4570460, 009-4570474, 009-4570489, 009-4570508, 009-4570510, DM550021, DM550024, DM550043, DM550051, DM550054. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Alcon Research LLC
Distributor
Worldwide - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI, PR and the countries of Canada, Japan, France.
Lot
UDI-DI: 00380650003997. Software Version: 1.5. Serial Numbers: 004-0680001, 004-0680002, 004-0680003, 004-0680004, 004-0680007, 004-0680008, 004-0680017, 004-0680026, 004-0680036, 004-0680047, 004-0680048, 004-0680086, 004-0680090, 004-0680092, 004-0680096, 004-0680100, 004-0680107, 004-0680112, 004-0680120, ,004-0680127,004-0680131, 004-0680134, 004-0680140, 004-0680152, 004-0680153, 004-0680161, 004-0680163, 004-0680188, 004-0680190,004-0680203, 004-0680206, 004-0680266, 004-0680267, 004-0680271, 004-0680273, 004-0680274, 004-0680283, 004-0680288, 004-0680293, 004-0680294, 004-0680306, 004-0680311, 004-0680315, 004-0680319, 004-0680325, 004-0680334, 004-0680336, 004-0680337, 004-0680338, 004-0680339, 004-0680346, 004-0680348, 004-0680352, 004-0680362, 004-0680404, 004-0680417, 004-0680418, 005-7230003, 005-7230005, 005-7230007, 005-7230017, 005-7230019, 005-7230021, 005-7230022, 005-7230023, 005-7230031, 005-7230035, 005-7230057, 005-7230081, 005-7230085, 005-7230086, 005-7230102, 005-7230108, 005-7230109, 005-7230186, 007-4610018, 007-4610023, 007-4610030, 007-4610032, 007-4610033, 007-4610034, 007-4610037, 007-4610044, 007-4610074, 007-4610107, 007-4610112, 007-4610115, 007-4610122, 007-4610123, 007-4610124, 007-4610125, 007-4610127, 007-4610132, 007-4610138, 007-4610166, 007-4610167, 007-4610179, 007-4610181, 007-4610186, 007-4610217, 007-4610219, 007-4610220, 007-4610226, 007-4610229, 007-4610230, 007-4610262, 007-4610275, 007-4610283, 007-4610287, 007-4610288, 007-4610308, 007-4610309, 007-4610351, 007-4610355, 007-4610359, 007-4610360, 007-4610366, 007-4610394, 007-4610404, 007-4610405, 007-4610406, 007-4610409, 009-4570001, 009-4570004, 009-4570005, 009-4570007, 009-4570013, 009-4570014, 009-4570015, 009-4570016, 009-4570019, 009-4570022, 009-4570033, 009-4570035, 009-4570044, 009-4570047, 009-4570048, 009-4570049, 009-4570051,009-4570052, 009-4570053, , 009-4570054,009-4570057, 009-4570059, 009-4570060, 009-4570064, 009-4570068, 009-4570070, 009-4570072, 009-4570073, 009-4570096, 009-4570097, 009-4570099, 009-4570100, 009-4570114, 009-4570116, 009-4570125, 009-4570127, 009-4570129, 009-4570130, 009-4570136, 009-4570137, 009-4570144, 009-4570148, 009-4570149, 009-4570150, 009-4570153, 009-4570155, 009-4570156, 009-4570158, 009-4570160, 009-4570161, 009-4570167, 009-4570171, 009-4570173, 009-4570176,009-4570178, 009-4570179, 009-4570180, 009-4570183, 009-4570184, 009-4570200, 009-4570204, 009-4570207, 009-4570233, 009-4570249, 009-4570256, 009-4570259, 009-4570263, 009-4570270, 009-4570275, 009-4570278, 009-4570281, 009-4570289, 009-4570293, 009-4570294, 009-4570300, 009-4570308, 009-4570317, 009-4570320, 009-4570331, 009-4570342, 009-4570343, 009-4570350, 009-4570353, 009-4570357, 009-4570365, 009-4570368, 009-4570393, 009-4570397, 009-4570425, 009-4570431, 009-4570435, 009-4570440, 009-4570448, 009-4570449, 009-4570451, 009-4570455, 009-4570457, 009-4570459, 009-4570460, 009-4570474, 009-4570489, 009-4570508, 009-4570510, DM550021, DM550024, DM550043, DM550051, DM550054
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399 recalled?

Yes. FDA published recall Z-2412-2024 on July 24, 2024. Reported hazard: After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.

What should I do if I have this product?

Voluntary: Firm initiated · Letter