FDAFDA · Food and drugs·Recall number Z-2416-2024

SlideView VS200 slide viewer,

Published August 7, 2024 · Recall initiated: July 22, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

What to do

FDA Mandated

Recall details

Manufacturer
EVIDENT SCIENTIFIC INC
Distributor
US
Lot
65159 67105 67104 69153 67106 69154
Risk classification
Class II
Area (FDA)
device
Country
United States