What's the hazard
A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.
What to do
FDA Mandated
Recall details
- Manufacturer
- EVIDENT SCIENTIFIC INC
- Distributor
- US
- Lot
- 65159 67105 67104 69153 67106 69154
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States