HYGENIC Wingless Fiesta Kit, Model no. H02790 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam.
Published July 1, 2020 · Recall initiated: June 3, 2020
Summary
FDA records medical device Class II recall Z-2422-2020, published July 1, 2020 and initiated June 3, 2020. Recalling firm: Coltene Whaledent Inc; Status: Ongoing; Reported quantity: 7,206 forceps; Lot: Batch numbers: J74546 J74184 If a distributor or end user received one of the batches listed above, they need to look at the batch number laser marked on the forceps themselves. If the forceps has any of the following batch numbers laser marked on them and the forceps do not fit with the dental dam clamps, the forceps should be returned to Coltene. Laser marked batch number: 1903 1904 1906 1907 1909 1910 1911 1915 1920 1921 1923 1927 1928 1930 1934 1940. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Coltene Whaledent Inc
- Distributor
- Pending
- Lot
- Batch numbers: J74546 J74184 If a distributor or end user received one of the batches listed above, they need to look at the batch number laser marked on the forceps themselves. If the forceps has any of the following batch numbers laser marked on them and the forceps do not fit with the dental dam clamps, the forceps should be returned to Coltene. Laser marked batch number: 1903 1904 1906 1907 1909 1910 1911 1915 1920 1921 1923 1927 1928 1930 1934 1940
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACEI Cutting Edge Instruments (PTC-57C)November 25, 2020
- FDAFG 57 100057X (100057X)November 25, 2020
- FDAHYGENIC Winged Fiesta Kit, Model no. H02778 - Product Usage: The dental dam forceps are intended to be used inJuly 1, 2020
- FDAAlly Bone Screw - Product Usage:Bone Screws are indicated for use in bone reconstruction, osteotomy,July 1, 2020
- FDAStumptown Cold Brew Coffee and Chocolate with Oatmilk. Product is TetraPak Prisma Pak aseptic packaging. KeepJuly 1, 2020
- FDALatex Dental Dams labeled as Non-Latex Dental DamsAugust 13, 2014
Questions about this recall
Was HYGENIC Wingless Fiesta Kit, Model no. H02790 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam. recalled?
Yes. FDA published recall Z-2422-2020 on July 1, 2020. Reported hazard: Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
What should I do if I have this product?
Voluntary: Firm initiated · Letter