Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable wound dressing that acts as an advanced wound care device. It is comprised of a non-woven pad entirely composed of HY AFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFFl 1 wound contact layer biodegradable matrix acts as a scaffold for cellular invasion and capillary growth. Hyalomatrix is indicated for the management of wounds including: partial and full-thickness wounds; second-degree bums; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.
Published July 18, 2018 · Recall initiated: May 10, 2018
Summary
FDA records medical device Class II recall Z-2423-2018, published July 18, 2018 and initiated May 10, 2018. Recalling firm: Anika Therapeutics, Inc.; Status: Terminated; Reported quantity: 1711; Lot: 2017017A, 2017033A, 2017034A, 2017031A, 2017065A, 2017066A, 2017007A, 2017008A, 2017030A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised.
What to do
Voluntary: Firm initiated · Other
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Anika Therapeutics, Inc.
- Distributor
- US Nationwide Distribution in the state of IL.
- Lot
- 2017017A, 2017033A, 2017034A, 2017031A, 2017065A, 2017066A, 2017007A, 2017008A, 2017030A
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAalba BOTANICA sport mineral sunscreen (titanium dioxide 7.0% and zinc oxide 9.0%), spf 45, 170 g (6 oz) tubesJuly 25, 2018
- FDAMethylcobalamin 1mg/mL Injection w/Pres. (Benzyl Alcohol 1% in Sterile Water), packaged in 10 mL and 30 mLJuly 25, 2018
- FDAPercepta Quad CRT-P MRI SureScan, REF W4TR04 Product Usage: The CRT-P models are implantable pulse generatorsJuly 18, 2018
- FDABioMerieux VIDAS System in combination with Quality Control VIDAS* (QCV*) Product Usage: the Quality ControlJuly 18, 2018
- FDAPercepta CRT-P MRI SureScan, REF W1TR01 Product Usage: The CRT-P models are implantable pulse generators withJuly 18, 2018
- FDAAnkle Arthrodesis Nail, right T2 Ankle ¿10x300mm, Product Number 18191030SJuly 18, 2018
Questions about this recall
Was Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable wound dressing that acts as an advanced wound care device. It is comprised of a non-woven pad entirely composed of HY AFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFFl 1 wound contact layer biodegradable matrix acts as a scaffold for cellular invasion and capillary growth. Hyalomatrix is indicated for the management of wounds including: partial and full-thickness wounds; second-degree bums; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use. recalled?
Yes. FDA published recall Z-2423-2018 on July 18, 2018. Reported hazard: Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised.
What should I do if I have this product?
Voluntary: Firm initiated · Other