Back to search
FDAFDA · Food and drugs·Recall number Z-2434-2024

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A

Published July 31, 2024 · Recall initiated: June 14, 2024

Summary

FDA records medical device Class II recall Z-2434-2024, published July 31, 2024 and initiated June 14, 2024. Recalling firm: B Braun Medical Inc; Status: Ongoing; Reported quantity: 9943; Lot: Lot Code: Model No 8713031U; UDI-DI 4046964797958; All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
B Braun Medical Inc
Distributor
Domestic distribution nationwide. International distribution to Canada.
Lot
Lot Code: Model No 8713031U; UDI-DI 4046964797958; All Serial Numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A recalled?

Yes. FDA published recall Z-2434-2024 on July 31, 2024. Reported hazard: Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

What should I do if I have this product?

Voluntary: Firm initiated · Letter