Terumo Advanced Perfusion System 1 ASSEMBLY FLOW POD Product Usage: The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.

Published October 3, 2012 · Recall initiated: September 10, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Terumo Cardiovascular Systems (CVS) received two reports of the false back flow alarms for the Terumo¿ Advanced Perfusion System 1 when there was no tubing in the flow sensor. Terumo CVS' investigation found that the voltage level on a circuit board was improperly set in a population of flow modules for Terumo System 1, the sensor operates correctly if the tubing is installed and it is highly unli

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide and US Nationwide Distribution
Lot
Catalog number: 802018 and serial numbers: 00032, 00042, 00056, 00147, 00623, 00700-00764, 00766-00820, and 00823-00847.
Risk classification
Class II
Area (FDA)
device
Country
United States