Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following descriptions: Bomba de infusi®n Plum 360 compatible con el software ICU Medical MedNet, Plum 360 Infusion System, PLUM 360, PLUM 360" Infuser, Plum 360 Infuser compatible with ICU Medical MedNet Software, Plum 360 Infuser compatible with ICU Medical MedNet Software Refurb, Plum 360 sistema infusionale compatibile con ICU Medical MedNet Software, Pompe a perfusion Plum 360, compatible avec le logiciel ICU Medical MedNet
Published August 21, 2024 · Recall initiated: May 7, 2024
Summary
FDA records medical device Class I recall Z-2447-2024, published August 21, 2024 and initiated May 7, 2024. Recalling firm: ICU Medical Inc; Status: Ongoing; Reported quantity: 11,961 units; Lot: 1. Product Code: 300100413. UDI Number (Serial Numbers): +M3353001004131 (43414527, 43414565); +M33530010041311 (43422532, 43511323, 43418697, 43419090, 43429221, 43419323, 43418389, 43418094, 43419257, 43429686, 43429767, 43429677, 43419283, 43429566, 43419102, 43429724, 43429718, 43429710, 43429707, 43429694, 43429692, 43429607, 43429593, 43429592, 43429582, 43429565, 43439515, 43418578, 43418655, 43484008, 43483343, 43483872, 43497023, 43496686, 43497667, 43497529, 43498447, 43496869, 43497174, 43496776, 43497353, 43498271, 43498371, 43496392, 43499242, 43497132, 43499181, 43427768, 43498473, 43496917, 43496871, 43496767, 43496675, 43497038, 43496717, 43498390, 43497559, 43497503, 43496747, 43496674, 43497662, 43497028, 43483719, 43497364, 43418579, 43418548, 43429728, 43418810, 43418646, 43418827, 43429624, 43418530, 43429743, 43429725, 43429563, 43429482, 43499274, 43496535, 43496740, 43483820, 43474669, 43477285, 43479179, 43478327, 43478398, 43497070, 43496861, 43496700, 43497350, 43483721, 43497377, 43497196, 43429558, 43429547, 43429534, 43429497, 43429494, 43429429, 43429399, 43429373, 43429371, 43429220, 43496723, 43429219, 43429067, 43493953, 43452640, 43498262, 43515328, 43496364, 43497508, 43497239, 43497605, 43497276, 43483973, 43484100, 43483749, 43483859, 43483951, 43429285, 43459787, 43429723, 43429679, 43429672, 43429602, 43429600, 43429571, 43429569, 43429560, 43429552, 43429478, 43429685, 43429678, 43497175, 43496779, 43497004, 43496661, 43496537, 43497568, 43429015, 43483722, 43483673, 43483819, 43483904, 43484184, 43483996, 43484136, 43484126, 43484062, 43483351, 43483929, 43495914, 43497555, 43450217, 43456114, 43429568, 43429562, 43429556, 43429555, 43429551, 43429525, 43429640, 43429595, 43429759, 43429758, 43429757, 43419195, 43429557, 43418513, 43429391, 43483901, 43483755, 43484065, 43483966, 43484021, 43483786, 43484034, 43483809, 43484043, 43429359, 43429393, 43429594, 43429733, 43429735, 43445560, 43457063, 43484047, 43483370, 43484038, 43483991, 43484087, 43483979, 43483886, 43483463, 43483967, 43419028, 43419169, 43483775, 30016791, 43476823, 43477643, 43477840, 43470631, 43477880, 43470383, 43470634, 43429333, 43429606, 43429752, 43429668, 43429731, 43429300, 43429580, 43429696, 43429543, 43429739, 43429705, 43429695, 43434798, 43445971, 43442858, 43480611, 43511708, 43458094, 43514602, 43475487, 43475851, 43476582, 43475914, 43476547, 43452416, 43453797, 43497890, 43495745, 43493977, 43466810, 43466861, 43499528, 43495859, 43459172, 43477297, 43475675, 43473381, 43478475, 43476749, 43471754, 43473403, 43473432, 43472746, 43479374, 43508275, 43499564, 43503906, 43477255, 43471276, 43471015, 43475938, 43469896, 43470540, 43470813, 43470906, 43469776, 43476238, 43478709, 43470128, 43477571, 43477247, 43471359, 43477641, 43477017, 43471375, 43499522, 43504108, 43456249, 30014556, 43494868, 43462861, 43467050, 43417069, 43502261, 43423943, 43424356, 43424211, 43417450, 43451631, 43459912, 43433440, 30017436, 43503961, 43490142, 30017827, 43458331, 43435039, 43476461, 43477865, 43479484, 43478395, 43454509, 43434374, 43476939, 43475322, 43476773, 43443855, 43491341, 43477718, 43476784, 43477689, 43477507, 43477026, 43475812, 43476512, 43477739, 43496664, 43496380, 43473890, 43475608, 43475479, 43475755, 43477582, 43477937, 43476747, 43477775, 43477673, 43424958, 43444105, 43466531, 43455873, 43491928, 43419088, 43509982, 43418662, 43419390, 43419337, 43419604, 43419435, 43419365, 43419387, 43419477, 43419382, 43419322, 43419366, 43445665, 43500995, 43428075, 43445371, 43494046, 43427707, 43504233, 43476692, 43477074, 43475700, 43470811, 43476663, 43476785, 43499516, 30016039, 30015138, 30015348, 30015672, 30015993, 30014687, 30015792, 30014884, 30014927, 30014854, 30016124, 43445597, 43453774, 43499719, 43496653, 43499462, 43494786, 43493852, 43495893, 43459272, 43484129, 43483337, 43483823, 43483366, 43483220, 43483338, 43483243, 43483880, 43483856, 43483815, 43499713, 43434862, 43443593, 4349716. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ICU Medical Inc
- Distributor
- Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay.
- Lot
- 1. Product Code: 300100413. UDI Number (Serial Numbers): +M3353001004131 (43414527, 43414565); +M33530010041311 (43422532, 43511323, 43418697, 43419090, 43429221, 43419323, 43418389, 43418094, 43419257, 43429686, 43429767, 43429677, 43419283, 43429566, 43419102, 43429724, 43429718, 43429710, 43429707, 43429694, 43429692, 43429607, 43429593, 43429592, 43429582, 43429565, 43439515, 43418578, 43418655, 43484008, 43483343, 43483872, 43497023, 43496686, 43497667, 43497529, 43498447, 43496869, 43497174, 43496776, 43497353, 43498271, 43498371, 43496392, 43499242, 43497132, 43499181, 43427768, 43498473, 43496917, 43496871, 43496767, 43496675, 43497038, 43496717, 43498390, 43497559, 43497503, 43496747, 43496674, 43497662, 43497028, 43483719, 43497364, 43418579, 43418548, 43429728, 43418810, 43418646, 43418827, 43429624, 43418530, 43429743, 43429725, 43429563, 43429482, 43499274, 43496535, 43496740, 43483820, 43474669, 43477285, 43479179, 43478327, 43478398, 43497070, 43496861, 43496700, 43497350, 43483721, 43497377, 43497196, 43429558, 43429547, 43429534, 43429497, 43429494, 43429429, 43429399, 43429373, 43429371, 43429220, 43496723, 43429219, 43429067, 43493953, 43452640, 43498262, 43515328, 43496364, 43497508, 43497239, 43497605, 43497276, 43483973, 43484100, 43483749, 43483859, 43483951, 43429285, 43459787, 43429723, 43429679, 43429672, 43429602, 43429600, 43429571, 43429569, 43429560, 43429552, 43429478, 43429685, 43429678, 43497175, 43496779, 43497004, 43496661, 43496537, 43497568, 43429015, 43483722, 43483673, 43483819, 43483904, 43484184, 43483996, 43484136, 43484126, 43484062, 43483351, 43483929, 43495914, 43497555, 43450217, 43456114, 43429568, 43429562, 43429556, 43429555, 43429551, 43429525, 43429640, 43429595, 43429759, 43429758, 43429757, 43419195, 43429557, 43418513, 43429391, 43483901, 43483755, 43484065, 43483966, 43484021, 43483786, 43484034, 43483809, 43484043, 43429359, 43429393, 43429594, 43429733, 43429735, 43445560, 43457063, 43484047, 43483370, 43484038, 43483991, 43484087, 43483979, 43483886, 43483463, 43483967, 43419028, 43419169, 43483775, 30016791, 43476823, 43477643, 43477840, 43470631, 43477880, 43470383, 43470634, 43429333, 43429606, 43429752, 43429668, 43429731, 43429300, 43429580, 43429696, 43429543, 43429739, 43429705, 43429695, 43434798, 43445971, 43442858, 43480611, 43511708, 43458094, 43514602, 43475487, 43475851, 43476582, 43475914, 43476547, 43452416, 43453797, 43497890, 43495745, 43493977, 43466810, 43466861, 43499528, 43495859, 43459172, 43477297, 43475675, 43473381, 43478475, 43476749, 43471754, 43473403, 43473432, 43472746, 43479374, 43508275, 43499564, 43503906, 43477255, 43471276, 43471015, 43475938, 43469896, 43470540, 43470813, 43470906, 43469776, 43476238, 43478709, 43470128, 43477571, 43477247, 43471359, 43477641, 43477017, 43471375, 43499522, 43504108, 43456249, 30014556, 43494868, 43462861, 43467050, 43417069, 43502261, 43423943, 43424356, 43424211, 43417450, 43451631, 43459912, 43433440, 30017436, 43503961, 43490142, 30017827, 43458331, 43435039, 43476461, 43477865, 43479484, 43478395, 43454509, 43434374, 43476939, 43475322, 43476773, 43443855, 43491341, 43477718, 43476784, 43477689, 43477507, 43477026, 43475812, 43476512, 43477739, 43496664, 43496380, 43473890, 43475608, 43475479, 43475755, 43477582, 43477937, 43476747, 43477775, 43477673, 43424958, 43444105, 43466531, 43455873, 43491928, 43419088, 43509982, 43418662, 43419390, 43419337, 43419604, 43419435, 43419365, 43419387, 43419477, 43419382, 43419322, 43419366, 43445665, 43500995, 43428075, 43445371, 43494046, 43427707, 43504233, 43476692, 43477074, 43475700, 43470811, 43476663, 43476785, 43499516, 30016039, 30015138, 30015348, 30015672, 30015993, 30014687, 30015792, 30014884, 30014927, 30014854, 30016124, 43445597, 43453774, 43499719, 43496653, 43499462, 43494786, 43493852, 43495893, 43459272, 43484129, 43483337, 43483823, 43483366, 43483220, 43483338, 43483243, 43483880, 43483856, 43483815, 43499713, 43434862, 43443593, 4349716
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPlum 360/A+ spare batteries. Item Number: SUB0000864.August 21, 2024
- FDAReplacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, PumpAugust 21, 2024
- FDAOrthopantomograph OP300 dental computed tomography x-ray system. The X-ray device produces conventional 2DAugust 21, 2024
- FDACold Smoked Capelin, 9-ounce packages, plastic trays, 34 packages per case (Best By Date:07.13.2024), 18August 21, 2024
- FDAASM Replacement Battery, Component Number SUB0000864May 24, 2023
- FDA0.9% Sodium Chloride Injection, USP, 100 mL flexible container bag, Rx Only, ICU Medical, Inc., Lake Forest,November 23, 2022
Questions about this recall
Was Replacement Battery List Number SUB0000864, found in Plum 360 Infusion System, List Number 30010. Product description is PLUM 360 Infuser. OUS units marketed with the following descriptions: Bomba de infusi®n Plum 360 compatible con el software ICU Medical MedNet, Plum 360 Infusion System, PLUM 360, PLUM 360" Infuser, Plum 360 Infuser compatible with ICU Medical MedNet Software, Plum 360 Infuser compatible with ICU Medical MedNet Software Refurb, Plum 360 sistema infusionale compatibile con ICU Medical MedNet Software, Pompe a perfusion Plum 360, compatible avec le logiciel ICU Medical MedNet recalled?
Yes. FDA published recall Z-2447-2024 on August 21, 2024. Reported hazard: Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.
What should I do if I have this product?
Voluntary: Firm initiated · Letter