Back to search
FDAFDA · Food and drugs·Recall number Z-2460-2012

Alere Triage TOX Drug Screen 9 Panel, P/N: 94402. The Alere Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses.

Published October 3, 2012 · Recall initiated: July 12, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Alere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in all meters.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Alere San Diego, Inc.
Distributor
Nationwide Distribution.
Lot
Lot Codes: K51245RB, W51022RB, W51207RB
Risk classification
Class II
Area (FDA)
device
Country
United States