FDAFDA · Food and drugs·Recall number Z-2468-2025

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUMBER UISH55V.

Published September 3, 2025 · Recall initiated: July 14, 2025

Summary

FDA records medical device Class II recall Z-2468-2025, published September 3, 2025 and initiated July 14, 2025. Recalling firm: American Contract Systems Inc; Status: Ongoing; Reported quantity: 204 units; Lot: 1) MODEL/ITEM NUMBER UICR10W: UDI/DI 00191072207241, Kit Lot Numbers: 42-7828711, 42-7980711, 42-8063211, 42-8124011; 2) MODEL/ITEM NUMBER UISH55V: UDI/DI 00191072235442, Kit Lot Numbers: 42-7799111, 42-7985211, 42-8065111, 42-8140711. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
American Contract Systems Inc
Distributor
US distribution to states of: IL and LA
Lot
1) MODEL/ITEM NUMBER UICR10W: UDI/DI 00191072207241, Kit Lot Numbers: 42-7828711, 42-7980711, 42-8063211, 42-8124011; 2) MODEL/ITEM NUMBER UISH55V: UDI/DI 00191072235442, Kit Lot Numbers: 42-7799111, 42-7985211, 42-8065111, 42-8140711.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) SFCH CRANI PACK (PS 132432), MODEL/ITEM NUMBER UICR10W; 2) SFCH SHUNT PACK (PS 139798), MODEL/ITEM NUMBER UISH55V. recalled?

Yes. FDA published recall Z-2468-2025 on September 3, 2025. Reported hazard: During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail