FDAFDA · Food and drugs·Recall number Z-2473-2023
Atellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma (lithium heparin Test Code: GGT Siemens Material Number (SMN): 11097597
Published September 6, 2023 · Recall initiated: July 10, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿CH Total Bilirubin_2 (TBil_2), Atellica¿ CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿ CH HDL Cholesterol (HDLC)
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam
- Lot
- Unique Device Identification (UDI): 00630414596440 All Lot Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States