FDAFDA · Food and drugs·Recall number Z-2476-2012
12/14 Neck Taper 00789503201 VerSys Femoral Head Provisional 32mm -3.5 00789503202 VerSys Femoral Head Provisional 32mm +0 00789503203 VerSys Femoral Head Provisional 32mm +3.5 00789503204 VerSys Femoral Head Provisional 32mm +7 00789503205 VerSys Femoral Head Provisional 32mm +10.5 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Published October 10, 2012 · Recall initiated: April 1, 2011
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
- Lot
- All lots distributed from lot 69998600 (2/20/1996) to present
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States