Back to search
FDAFDA · Food and drugs·Recall number Z-2478-2025

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Published September 17, 2025 · Recall initiated: August 4, 2025

Summary

FDA records medical device Class II recall Z-2478-2025, published September 17, 2025 and initiated August 4, 2025. Recalling firm: WASATCH PHOTONICS; Status: Ongoing; Reported quantity: 192; Lot: all codes included in the model numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

What to do

FDA Mandated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
WASATCH PHOTONICS
Distributor
U.S.
Lot
all codes included in the model numbers.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC. recalled?

Yes. FDA published recall Z-2478-2025 on September 17, 2025. Reported hazard: Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

What should I do if I have this product?

FDA Mandated