Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Published September 17, 2025 · Recall initiated: August 4, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

What to do

FDA Mandated

Recall details

Manufacturer
WASATCH PHOTONICS
Distributor
U.S.
Lot
all codes included in the model numbers.
Risk classification
Class II
Area (FDA)
device
Country
United States