Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.
Published September 17, 2025 · Recall initiated: August 4, 2025
Summary
FDA records medical device Class II recall Z-2478-2025, published September 17, 2025 and initiated August 4, 2025. Recalling firm: WASATCH PHOTONICS; Status: Ongoing; Reported quantity: 192; Lot: all codes included in the model numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
What to do
FDA Mandated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- WASATCH PHOTONICS
- Distributor
- U.S.
- Lot
- all codes included in the model numbers.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHershey s Ice Cream, Green Mint Chip (GMC) Bulk, 3 gallons, 1 unit , bulk can- a cardboard tub with metalSeptember 17, 2025
- FDAArtis Pheno. Image-Intensified Flouroscopic X-Ray System.September 17, 2025
- FDAPeanut rounds 2.25 oz and 3 ozSeptember 17, 2025
- FDABoston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)September 17, 2025
- FDAOrganic BABY bedtime drops; sleep + immunity blend; Promotes restful sleep with natural herbs & supportsSeptember 17, 2025
- FDAIntelliSpace Cardiovascular, Software 8.0.0.4.September 17, 2025
Questions about this recall
Was Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC. recalled?
Yes. FDA published recall Z-2478-2025 on September 17, 2025. Reported hazard: Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
What should I do if I have this product?
FDA Mandated