Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.
Published October 10, 2012 · Recall initiated: August 23, 2012
Summary
FDA records medical device Class II recall Z-2485-2012, published October 10, 2012 and initiated August 23, 2012. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 268 units total; Lot: Catalog number: 14-400795; and lot numbers: 133420, 409460, 469960, 487650, 487660, 627940, 627950, 797190, and 797200. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of AL, GA, MT, TA, AZ, WA, CA, TA, IN, NY, NJ, VA, MA, MI, MD, PA, KY, NC, UT, NH, FL, OK, CO, LA, OH, MO and SD, and the countries of Japan, Netherlands, Australia, and Costa Rica.
- Lot
- Catalog number: 14-400795; and lot numbers: 133420, 409460, 469960, 487650, 487660, 627940, 627950, 797190, and 797200.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDA00780304001 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780304002 MIS 12/14 Slotted Femoral HeadOctober 10, 2012
- FDAOptilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is intended forOctober 10, 2012
Questions about this recall
Was Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone. recalled?
Yes. FDA published recall Z-2485-2012 on October 10, 2012. Reported hazard: Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw
What should I do if I have this product?
Voluntary: Firm initiated · Letter