Universal Pack II (2) Customed 900-2509 (1) MAYO STAND COVER REINFORCED (2) DRAPE SIDE 42" X 76" WITH TAPE (1) DRAPE BOTTOM ABSORBENT REINFORCED 71" X 62" STD SMS WITH TAPE (1) DRAPE TOP W/ADHESIVE 1 08" X 50" STD SMS (1) SUTURE BAG FLORAL (1) TABLE COVER REINFORCED 50" X 90" (1) GOWN SURG. REINFORCED XLGE TOWEL I WRAP Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Published September 10, 2014 · Recall initiated: May 20, 2014
Summary
FDA records medical device Class I recall Z-2485-2014, published September 10, 2014 and initiated May 20, 2014. Recalling firm: Customed, Inc; Status: Terminated; Reported quantity: 3 lots, 612 units, multiple units per pack; Lot: 900-2509, 3 lots: 112020357 112051971 113078530. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Customed, Inc
- Distributor
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot
- 900-2509, 3 lots: 112020357 112051971 113078530
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACatalog Number: 9002615 EXTREMITY SURGICAL PACK Used by medical staff to perform surgical procedures andApril 20, 2016
- FDACatalog Number: 9001482 ENT SURGICAL PACK Used by medical staff to perform surgical procedures and patientApril 20, 2016
- FDACatalog Number: 9003190 SURG GOWN SMS AMMI IV FULL Used by medical staff to perform surgical procedures andApril 20, 2016
- FDACatalog Number: 900110 TRACHEOSTOMY CARE KIT WITH TRACH TUBE HOLDER FOR ADULT Used by medical staff to performApril 20, 2016
- FDAHIP PACK I (2), CUSTOMED, code 900-2504 (1) GOWN SURG. REINFORCED XLGE TOWEL/WRAP (1) MAYO STAND COVERSeptember 10, 2014
- FDAShoulder pack (I) (2), code 900-2515 (1) GOWN IMP. XTRA REINF. XL TOWEL WRAP LEVEL Ill (1) TOWEL ABSORBENT 15"September 10, 2014
Questions about this recall
Was Universal Pack II (2) Customed 900-2509 (1) MAYO STAND COVER REINFORCED (2) DRAPE SIDE 42" X 76" WITH TAPE (1) DRAPE BOTTOM ABSORBENT REINFORCED 71" X 62" STD SMS WITH TAPE (1) DRAPE TOP W/ADHESIVE 1 08" X 50" STD SMS (1) SUTURE BAG FLORAL (1) TABLE COVER REINFORCED 50" X 90" (1) GOWN SURG. REINFORCED XLGE TOWEL I WRAP Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents. recalled?
Yes. FDA published recall Z-2485-2014 on September 10, 2014. Reported hazard: Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
What should I do if I have this product?
Voluntary: Firm initiated · Letter