FDAFDA · Food and drugs·Recall number Z-2486-2026

MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACK, Medline Kit/SKU DYNJ67331B; 2) CARDIAC SURGERY, Medline Kit/SKU DYNJ911589A; 3) CAROTID ENDARTECOMY, Medline Kit/SKU DYNJ36211D; 4) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973; 5) CHLA ECMO / CARDIAC PACK, Medline Kit/SKU DYNJ87973A; 6) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879; 7) CV PART 1 AND 2, Medline Kit/SKU DYNJ908879C; 8) FEM POP CDS, Medline Kit/SKU CDS983670K; 9) LAP BASIN TRAY, Medline Kit/SKU DYNJ50908B; 10) OPEN HEART LTX CATHETERS KIT, Medline Kit/SKU DYNJ0049032A; 11) OPEN HEART PACK, Medline Kit/SKU DYNJ80549C; 12) OPEN HEART PACK, Medline Kit/SKU DYNJ80549D; 13) RR-ADULT ECMO PACK, Medline Kit/SKU DYNJ69742B; 14) TAVR, Medline Kit/SKU DYNJ906509C.

Published June 24, 2026 · Recall initiated: May 4, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Medline Industries, LP
Distributor
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Lot
Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 23JBM539; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 23IBF088; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 23FBM203; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22JBK233; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22JBB115; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22IBO304; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22HBR331; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22FBM460; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22DBS093; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22CBJ072; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 22BBX288; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 21LBU240; Medline Kit Number/SKU DYNJ67331B: UDI/DI 10195327013837 (each), 40195327013838 (case), Lot Number 21JBP452; Medline Kit Number/SKU DYNJ911589A: UDI/DI 10198459646102 (each), 40198459646103 (case), Lot Number 26CBD487; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 24EBB811; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 24ABV035; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 23LBG394; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 23KBA664; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 23HBC987; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 23GBQ638; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 23EBM118; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 23DBB008; Medline Kit Number/SKU DYNJ36211D: UDI/DI 10889942885657 (each), 40889942885658 (case), Lot Number 22KBR412; Medline Kit Number/SKU DYNJ87973: UDI/DI 10198459085031 (each), 40198459085032 (case), Lot Number 24HMH646; Medline Kit Number/SKU DYNJ87973A: UDI/DI 10198459159633 (each), 40198459159634 (case), Lot Number 24JMG070; Medline Kit Number/SKU DYNJ908879: UDI/DI 10195327203429 (each), 40195327203420 (case), Lot Number 22HBX489; Medline Kit Number/SKU DYNJ908879C: UDI/DI 10195327690885 (each), 40195327690886 (case), Lot Number 24KBD462; Medline Kit Number/SKU DYNJ908879C: UDI/DI 10195327690885 (each), 40195327690886 (case), Lot Number 24IBQ955; Medline Kit Number/SKU DYNJ908879C: UDI/DI 10195327690885 (each), 40195327690886 (case), Lot Number 24GBA028; Medline Kit Number/SKU DYNJ908879C: UDI/DI 10195327690885 (each), 40195327690886 (case), Lot Number 24FBI615; Medline Kit Number/SKU CDS983670K: UDI/DI 10193489952568 (each), 40193489952569 (case), Lot Number 26BBJ555; Medline Kit Number/SKU CDS983670K: UDI/DI 10193489952568 (each), 40193489952569 (case), Lot Number 26BBB853; Medline Kit Number/SKU CDS983670K: UDI/DI 10193489952568 (each), 40193489952569 (case), Lot Number 25LBA306; Medline Kit Number/SKU CDS983670K: UDI/DI 10193489952568 (each), 40193489952569 (case), Lot Number 25IBD419; Medline Kit Number/SKU CDS983670K: UDI/DI 10193489952568 (each), 40193489952569 (case), Lot Number 25HBM069; Medline Kit Number/SKU CDS983670K: UDI/DI 10193489952568 (each), 40193489952569 (case), Lot Number 25GBT806;
Risk classification
Class II
Area (FDA)
device
Country
United States