Back to search
FDAFDA · Food and drugs·Recall number Z-2487-2024

BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow, REF: 169-131 containing software version 3.9.1.9

Published August 14, 2024 · Recall initiated: June 24, 2024

Summary

FDA records medical device Class II recall Z-2487-2024, published August 14, 2024 and initiated June 24, 2024. Recalling firm: CareFusion 303, Inc.; Status: Ongoing; Reported quantity: 61 units; Lot: UDI: 10885403512568/ Serial Number: 16139674 16124908 16063924 16084467 16221549 16139675 16220875 16237258 16284001 15940400 15940104 15941703 15941654 15943750 15943749 15500696 15654711 15654721 15631099 15631104 15775749 15383480 15383481 15383475 15383486 15383487 15366328 15606170 15575974 15575976 15490745 15422133 15853047 15902413 15598095 15550853 15437809 15437812 15518269 15518273 15518270 15872432 15872429 15662816 15504833 15504836 15503511 15430948 15430963 15639687 15639724 15639720 15639718 15328493 15442794 15442789 15761398 15761401 15937579 15786154 15684803. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
CareFusion 303, Inc.
Distributor
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Lot
UDI: 10885403512568/ Serial Number: 16139674 16124908 16063924 16084467 16221549 16139675 16220875 16237258 16284001 15940400 15940104 15941703 15941654 15943750 15943749 15500696 15654711 15654721 15631099 15631104 15775749 15383480 15383481 15383475 15383486 15383487 15366328 15606170 15575974 15575976 15490745 15422133 15853047 15902413 15598095 15550853 15437809 15437812 15518269 15518273 15518270 15872432 15872429 15662816 15504833 15504836 15503511 15430948 15430963 15639687 15639724 15639720 15639718 15328493 15442794 15442789 15761398 15761401 15937579 15786154 15684803
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was BD PYXIS MDBNK TWR MN CR-4HH-5FM-P RxNow, REF: 169-131 containing software version 3.9.1.9 recalled?

Yes. FDA published recall Z-2487-2024 on August 14, 2024. Reported hazard: Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.

What should I do if I have this product?

Voluntary: Firm initiated · Letter