FDAFDA · Food and drugs·Recall number Z-2493-2018

Hydroline Trumpet Valve, Trumpet Valve, 5mm x 28cm Probe, Double Spike, Ref ASU1211, packaged 1/box, 12 boxes/case, single use, RX, Sterile. Manufactured For Cardinal Health, Waukegan, IL.

Published July 25, 2018 · Recall initiated: June 11, 2018

Summary

FDA records medical device Class II recall Z-2493-2018, published July 25, 2018 and initiated June 11, 2018. Recalling firm: Cardinal Health 200, LLC; Status: Terminated; Reported quantity: A total of 63,020 valves were distributed. Specific product quantities were not provided.; Lot: Lot number 41736. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The suction valve may not close properly which could cause continuous suction.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cardinal Health 200, LLC
Distributor
Distribution was made to AL, AR, AZ, CA, CO, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MT, NC, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, and WY. Government and military distribution was also made. Foreign distribution was made to Belgium, France, Germany, Israel, Italy, Netherlands, Singapore, and United Kingdom.
Lot
Lot number 41736.
Risk classification
Class II
Area (FDA)
device
Country
United States