FDAFDA · Food and drugs·Recall number Z-2496-2020
TOMTEC-ARENA TTA2;software version no. 2.20 and lower; Picture archiving and communications system. Software Versions and applicable UDI: TTA 1.2 release prior to UDI requirement TTA2.00.01 release prior to UDI requirement TTA2.11.01 release prior to UDI requirement TTA2.11.02 release prior to UDI requirement TTA2.20.01 UDI: (01)04260361440143(11)160429(10)20.01 TTA2.20.10 UDI:(01)04260361440143(11)160809(10)20.10 TTA2.20.13 UDI:(01)04260361440143(11)160922(10)20.13 TTA2.20.14 UDI:(01)04260361440143(11)161028(10)20.14 TTA2.20.15 UDI: (01)04260361440143[11]161207[10]20.15 TTA2.20.16 UDI:(01)04260361440143(11)170127(10)20.16 TTA2.20.18 UDI:(01)04260361440143(11)170622(10)20.18
Published July 15, 2020 · Recall initiated: September 23, 2019
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm discovered a software issue associated with the Image-Com module/clinical application package (CAP) interface for 3D application in the ARENA software, where measurements from one study might be stored to another study.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Tomtec Imaging Systems Gmbh
- Distributor
- US:AL, AZ, CA, CT, IL, MA, MI,NC, NJ, NY, OH, PA, RI, TX, VA. OUS: Australia, Canada, Czech Republic, France, Germany, Italy, Latvia, Sweden, Switzerland, United Kingdom.
- Lot
- The affected product versions are TOMTEC-ARENA, TTA 2.20 and lower, with modules 4D LV-Analysis, 4D RV-Function, 4D MV-Assessment, 3D Cardio-View, 4D Sono-Scan, Echo-Com, 2D Cardiac Performance Analysis, and QAngio.
- Risk classification
- Class II