BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P, REF: 169-96 containing software version 3.9.1.9
Published August 14, 2024 · Recall initiated: June 24, 2024
Summary
FDA records medical device Class II recall Z-2500-2024, published August 14, 2024 and initiated June 24, 2024. Recalling firm: CareFusion 303, Inc.; Status: Ongoing; Reported quantity: 1 unit; Lot: UDI: 10885403512568/ Serial Number: 15684798. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV.
- Lot
- UDI: 10885403512568/ Serial Number: 15684798
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Pyxis Anesthesia Station 4000 REF: 338 Medication cabinetJune 10, 2026
- FDABD Pyxis MedBank Tower REF: 1145-00 Medication cabinetJune 10, 2026
- FDABD Pyxis Mini Main REF: 349 Medication cabinetJune 10, 2026
- FDABD Pyxis MedStation ES REF: 323 Medication cabinetJune 10, 2026
- FDABD PYXIS MEDBANK TWR MN CR-6HH-4FM-P, REF: 169-94 containing software version 3.9.1.9August 14, 2024
- FDABD PYXIS MEDBANK TWR MN CR-8HH-2FH-1FM-P, REF: 169-97 containing software version 3.9.1.9August 14, 2024
Questions about this recall
Was BD PYXIS MEDBANK TWR MN CR-8HH-1FH-2FM-P, REF: 169-96 containing software version 3.9.1.9 recalled?
Yes. FDA published recall Z-2500-2024 on August 14, 2024. Reported hazard: Due to software issue, there is a potential when a restock label is printed for a medication stored in automated dispensing cabinets, that the incorrect bin for the medication is printed on the label.
What should I do if I have this product?
Voluntary: Firm initiated · Letter