Back to search
FDAFDA · Food and drugs·Recall number Z-2502-2026

MEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B

Published June 24, 2026 · Recall initiated: May 4, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Medline Industries, LP
Distributor
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Lot
UDI/DI 10195327548261 (each), 40195327548262 (case). Lot Numbers: 25LBI062, 25HBR967, 24JBK618, 24BBQ860.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card