IntelliVue Patient Monitor MX450 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866062. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates.
Published August 14, 2024 · Recall initiated: June 28, 2024
Summary
FDA records medical device Class II recall Z-2509-2024, published August 14, 2024 and initiated June 28, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 49 units; Lot: UDI: 00884838038769. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Nationwide Foreign: Australia Austria Bangladesh Brazil Canada Chile China Colombia Denmark France Germany Hong Kong India Indonesia Italy Japan Netherlands Portugal Singapore South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand United Kingdom Vietnam
- Lot
- UDI: 00884838038769
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDAIntelliVue Patient Monitor MX400 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866060.August 14, 2024
- FDAIntelliVue Patient Monitor MX500 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866064.August 14, 2024
Questions about this recall
Was IntelliVue Patient Monitor MX450 with L.x and M.x; K.x provided with L.x entitlements Model Number: 866062. Indicated for use by health care professionals for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. recalled?
Yes. FDA published recall Z-2509-2024 on August 14, 2024. Reported hazard: Philips lntelliVue Patient Monitors MX400/450/500/550 Shipped or Upgraded with Incorrect Software Options. Changes were made to reflect some options becoming standard capabilities for software-version N.x. In the process, the entitlements of software options MOS, M06, and M20 were removed for software versions L.x and M.x. As a result of these changes Patient Monitors manufactured with or updated to the latest versions of software L.x or M.x will not offer the capabilities> Mmonitors with software version K.x may experience this issue too, due to software version K being out of support and the devices being provided with L.x entitlements offered by software options MOS, M06, and M20. Potential for incorrect or delayed treatment for the patient,
What should I do if I have this product?
Voluntary: Firm initiated · Letter