FDAFDA · Food and drugs·Recall number Z-2510-2015

NovaPlus Trach Care* Closed Suction System for Adults, Double Swivel Elbow, Directional Tip, 14F (Product Code 2V22109) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.

Published September 9, 2015 · Recall initiated: July 29, 2015

Summary

FDA records medical device Class II recall Z-2510-2015, published September 9, 2015 and initiated July 29, 2015. Recalling firm: Halyard Health, Inc; Status: Terminated; Reported quantity: N/A; Lot: Product Code 2V22109 - M5076T629. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Halyard Health, Inc
Distributor
Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada
Lot
Product Code 2V22109 - M5076T629
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was NovaPlus Trach Care* Closed Suction System for Adults, Double Swivel Elbow, Directional Tip, 14F (Product Code 2V22109) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway. recalled?

Yes. FDA published recall Z-2510-2015 on September 9, 2015. Reported hazard: The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

What should I do if I have this product?

Voluntary: Firm initiated · Letter