FDAFDA · Food and drugs·Recall number Z-2514-2024
BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.
Published August 14, 2024 · Recall initiated: May 16, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- BioFire Diagnostics, LLC
- Distributor
- U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.
- Lot
- Part Number: RFIT-ASY-0116 (30 test kit) UDI-DI code: 00815381020109 0081538102011: All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Part Number: RFIT-ASY-0104 (6 test kit) UDI-DI code: 0081538102011 All unexpired products are potentially impacted if used in conjunction with the Remel Cary Blair transport medium Remel Cary-Blair Transport Medium lot numbers: 743930 769240 712258 782425 732357 742049
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States