VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
Published July 1, 2026 · Recall initiated: May 8, 2026
Summary
FDA records medical device Class II recall Z-2525-2026, published July 1, 2026 and initiated May 8, 2026. Recalling firm: Ventec Life Systems, Inc.; Status: Ongoing; Reported quantity: 1; Lot: REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ventec Life Systems, Inc.
- Distributor
- US Nationwide distribution in the state of UT.
- Lot
- REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.October 28, 2026
- FDAVC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.October 28, 2026
- FDAVCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.October 28, 2026
- FDAVOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.October 28, 2026
- FDATMJ Bilateral Implants, REF: CHG020July 1, 2026
- FDAMulti-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310July 1, 2026
Questions about this recall
Was VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001 recalled?
Yes. FDA published recall Z-2525-2026 on July 1, 2026. Reported hazard: Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
What should I do if I have this product?
Voluntary: Firm initiated · Letter