VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Published July 1, 2026 · Recall initiated: May 8, 2026

Summary

FDA records medical device Class II recall Z-2525-2026, published July 1, 2026 and initiated May 8, 2026. Recalling firm: Ventec Life Systems, Inc.; Status: Ongoing; Reported quantity: 1; Lot: REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ventec Life Systems, Inc.
Distributor
US Nationwide distribution in the state of UT.
Lot
REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001 recalled?

Yes. FDA published recall Z-2525-2026 on July 1, 2026. Reported hazard: Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

What should I do if I have this product?

Voluntary: Firm initiated · Letter