FDAFDA · Food and drugs·Recall number Z-2528-2014
Laparoscopy pack, , Code 900-3178, contains: (1) BULB SYRINGE 60cc LIF (10) GAUZE SPONGE 4" X 4" 16PLY XRD LIF (1) MAYO STAND COVER REINFORCED LIF (1) TABLE COVER REINFORCED 50" X 90" LIF (3) GOWN SURG REINFORCED XL EXTRA LONG LEVEL IV (4) TOWEL ABSORBENT 15" X 20" LIF (1) DRAPE VIDEO CAMERA 13cm X 244cm LIF (3) LITE GLOVE LIF (1) DRAPE LAP ABDOMINAL W/POUCH 102" X 122" X 78" (1) BLADE SURGICAL# 11 (1) SYRINGE 10cc WITHOUT NEEDLE LUER LOCK LIF (4) DRAPE UTILITY WITH TAPE LIF (1) SUTURE BAG FLORAL L/F (1) NEEDLE & BLADE COUNTER 10c MAG/CLEAR LIF (1) CABLE LAP 10FR 4MM MALE MONOP. LIF (1) SPECIMEN CONTAINER 4oz W/LID & LABEL (2) SHEET% 60" X 77" DRAPE REINFORCED LIF (1) TUBING INSUFFLATION SET W/0 RING ADAPTOR LIF (1) WRAPPER 24" X 24" LIF (1) ULTRA VERES NEEDLE 120MM LIF (1) TROCAR ENDO.XCELL DILATING TIP 11 MM (1) TROCAR ENDO.XCELL DILATING TIP 5MM LIF (1) CANNULA 5MM XCEL SLEEVES LIF (2) DENTURE CUP 8oz W/LID (2) DRESSING NON ADH TELFA 4" X 3" (1) CHLORAPREP 26ML APPLICATOR TEAL LIF Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Published September 10, 2014 · Recall initiated: May 20, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Customed, Inc
- Distributor
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot
- 900-3178, 4 lots: 131210899 140211837 140312152 140513143
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States