Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Published September 27, 2023 · Recall initiated: July 26, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hamilton Medical, Inc.
Distributor
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Lot
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Risk classification
Class I
Area (FDA)
device
Country
United States