Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator

Published September 27, 2023 · Recall initiated: July 26, 2023

Summary

FDA records medical device Class I recall Z-2529-2023, published September 27, 2023 and initiated July 26, 2023. Recalling firm: Hamilton Medical, Inc.; Status: Ongoing; Reported quantity: 6 units; Lot: UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hamilton Medical, Inc.
Distributor
Distribution US nationwide, including Puerto Rico. OUS: Not provided
Lot
UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator recalled?

Yes. FDA published recall Z-2529-2023 on September 27, 2023. Reported hazard: Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

What should I do if I have this product?

Voluntary: Firm initiated · Letter