Back to search
FDAFDA · Food and drugs·Recall number Z-2529-2026

Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040242

Published July 1, 2026 · Recall initiated: May 20, 2026

Summary

FDA records medical device Class II recall Z-2529-2026, published July 1, 2026 and initiated May 20, 2026. Recalling firm: Zimmer, Inc.; Status: Ongoing; Reported quantity: 429 units; Lot: Lot Code: Model No. 110040242 ; UDI-DI (01)00889024681989(17)350601(10)099471 ; Lot Number 099471 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350614(10)099473 ; Lot Number 099473 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350804(10)099474 ; Lot Number 099474 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350806(10)099475 ; Lot Number 099475. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.

What to do

Voluntary: Firm initiated

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
US Nationwide distribution.
Lot
Lot Code: Model No. 110040242 ; UDI-DI (01)00889024681989(17)350601(10)099471 ; Lot Number 099471 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350614(10)099473 ; Lot Number 099473 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350804(10)099474 ; Lot Number 099474 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350806(10)099475 ; Lot Number 099475
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040242 recalled?

Yes. FDA published recall Z-2529-2026 on July 1, 2026. Reported hazard: Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.

What should I do if I have this product?

Voluntary: Firm initiated