FDAFDA · Food and drugs·Recall number Z-2539-2020

OrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - Product Usage: is intended for fixation of tendons and ligaments during orthopedic reconstruction procedures such as Anterior Cruciate Ligament (ACL) Reconstruction.

Published July 15, 2020 · Recall initiated: June 5, 2020

Summary

FDA records medical device Class II recall Z-2539-2020, published July 15, 2020 and initiated June 5, 2020. Recalling firm: OrthoPediatrics Corp; Status: Terminated; Reported quantity: 11 units; Lot: UDI: (01)00841132122211(17)240722(10)214320319A02 Lot#: 214320319A02. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was found to have 10-1008-4060 product in the package. Product labeled as 10-1008-4060 was found to have 10-1008-4070 product in the package.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
OrthoPediatrics Corp
Distributor
US Nationwide distribution.
Lot
UDI: (01)00841132122211(17)240722(10)214320319A02 Lot#: 214320319A02
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - Product Usage: is intended for fixation of tendons and ligaments during orthopedic reconstruction procedures such as Anterior Cruciate Ligament (ACL) Reconstruction. recalled?

Yes. FDA published recall Z-2539-2020 on July 15, 2020. Reported hazard: ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was found to have 10-1008-4060 product in the package. Product labeled as 10-1008-4060 was found to have 10-1008-4070 product in the package.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail