FDAFDA · Food and drugs·Recall number Z-2548-2014

Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA

Published September 3, 2014 · Recall initiated: July 22, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution.
Lot
Item Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500 Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644
Risk classification
Class II
Area (FDA)
device
Country
United States