Monaco RTP System, 5.50/5.51, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review

Published October 2, 2019 · Recall initiated: September 4, 2019

Summary

FDA records medical device Class II recall Z-2561-2019, published October 2, 2019 and initiated September 4, 2019. Recalling firm: Elekta Inc; Status: Ongoing; Reported quantity: 49 units; Lot: UDI: (01)00858164002244(10) 5.50.00 (01)00858164002268(10) 5.51.00. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Elekta Inc
Distributor
Worldwide Distribution - US Nationwide in the states of WI, OH, NY, OH, IN, IA, TX, and countries of Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, UK
Lot
UDI: (01)00858164002244(10) 5.50.00 (01)00858164002268(10) 5.51.00
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Monaco RTP System, 5.50/5.51, radiation treatment planning system Product Usage: The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and carbon treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for: -contouring -image manipulation -simulation -image fusion -plan optimization -QA and plan review recalled?

Yes. FDA published recall Z-2561-2019 on October 2, 2019. Reported hazard: It is possible that the forced electron density settings will be changed for some structures unintentionally and this can result in incorrect dose calculation.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail