Back to search
FDAFDA · Food and drugs·Recall number Z-2565-2018

Tearaway Introducers, Catalog #: a) BNS-000-07, b) BNS-000-10; c) DIL-015-50; d) DSS-005-11; e) DSS-005-12; f) DSS-006-045; g) DSS-007-05; h) DSS-007-165; i) DSS-010-03; j) DSS-010-04; k) DSS-010-045; l) DSS-010-05; m) DSS-012-07; n) SPT-003-46

Published August 8, 2018 · Recall initiated: May 2, 2018

Summary

FDA records medical device Class II recall Z-2565-2018, published August 8, 2018 and initiated May 2, 2018. Recalling firm: Galt Medical Corporation; Status: Terminated; Reported quantity: 23777; Lot: Lot #: a) 18024880, 18043086, 18066833, 18073282; b) 18039555, 18059249, 18074835; c) 17363134; d) 17352818, 18051601; e) 18019228, 18047509; f) 18023833; g) 17356436, 18018694, 18018695, 18040074; h) 18061579; i) 18015027; j) 18019038, 18029067, 18033068, 18075816, 18075817, 18075818; k) 17363816, 18023834, 18026040, 18026072, 18033073, 18033079, 18054248; l) 18005018, 18012032, 18012034, 18012035, 18012036, 18019037, 18026953, 18089355, 18096357, 18096358, 18103359; m) 18047202; n) 18054349. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Galt Medical Corporation
Distributor
U.S.: MN, PA, VA, CA, UT, FL, NY, GA, LA, MO, TX, MS, OH, WA, MA, IL, KY, NC, and Puerto Rico; OUS: Canada, Australia, France, Germany, Ireland, Italy, Netherlands, Poland, Slovakia, Switzerland, Belgium, United Kingdom, Korea
Lot
Lot #: a) 18024880, 18043086, 18066833, 18073282; b) 18039555, 18059249, 18074835; c) 17363134; d) 17352818, 18051601; e) 18019228, 18047509; f) 18023833; g) 17356436, 18018694, 18018695, 18040074; h) 18061579; i) 18015027; j) 18019038, 18029067, 18033068, 18075816, 18075817, 18075818; k) 17363816, 18023834, 18026040, 18026072, 18033073, 18033079, 18054248; l) 18005018, 18012032, 18012034, 18012035, 18012036, 18019037, 18026953, 18089355, 18096357, 18096358, 18103359; m) 18047202; n) 18054349
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Tearaway Introducers, Catalog #: a) BNS-000-07, b) BNS-000-10; c) DIL-015-50; d) DSS-005-11; e) DSS-005-12; f) DSS-006-045; g) DSS-007-05; h) DSS-007-165; i) DSS-010-03; j) DSS-010-04; k) DSS-010-045; l) DSS-010-05; m) DSS-012-07; n) SPT-003-46 recalled?

Yes. FDA published recall Z-2565-2018 on August 8, 2018. Reported hazard: The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter