Tearaway Introducers, Catalog #: a) BNS-000-07, b) BNS-000-10; c) DIL-015-50; d) DSS-005-11; e) DSS-005-12; f) DSS-006-045; g) DSS-007-05; h) DSS-007-165; i) DSS-010-03; j) DSS-010-04; k) DSS-010-045; l) DSS-010-05; m) DSS-012-07; n) SPT-003-46
Published August 8, 2018 · Recall initiated: May 2, 2018
Summary
FDA records medical device Class II recall Z-2565-2018, published August 8, 2018 and initiated May 2, 2018. Recalling firm: Galt Medical Corporation; Status: Terminated; Reported quantity: 23777; Lot: Lot #: a) 18024880, 18043086, 18066833, 18073282; b) 18039555, 18059249, 18074835; c) 17363134; d) 17352818, 18051601; e) 18019228, 18047509; f) 18023833; g) 17356436, 18018694, 18018695, 18040074; h) 18061579; i) 18015027; j) 18019038, 18029067, 18033068, 18075816, 18075817, 18075818; k) 17363816, 18023834, 18026040, 18026072, 18033073, 18033079, 18054248; l) 18005018, 18012032, 18012034, 18012035, 18012036, 18019037, 18026953, 18089355, 18096357, 18096358, 18103359; m) 18047202; n) 18054349. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Galt Medical Corporation
- Distributor
- U.S.: MN, PA, VA, CA, UT, FL, NY, GA, LA, MO, TX, MS, OH, WA, MA, IL, KY, NC, and Puerto Rico; OUS: Canada, Australia, France, Germany, Ireland, Italy, Netherlands, Poland, Slovakia, Switzerland, Belgium, United Kingdom, Korea
- Lot
- Lot #: a) 18024880, 18043086, 18066833, 18073282; b) 18039555, 18059249, 18074835; c) 17363134; d) 17352818, 18051601; e) 18019228, 18047509; f) 18023833; g) 17356436, 18018694, 18018695, 18040074; h) 18061579; i) 18015027; j) 18019038, 18029067, 18033068, 18075816, 18075817, 18075818; k) 17363816, 18023834, 18026040, 18026072, 18033073, 18033079, 18054248; l) 18005018, 18012032, 18012034, 18012035, 18012036, 18019037, 18026953, 18089355, 18096357, 18096358, 18103359; m) 18047202; n) 18054349
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABoston Scientific Enhance Transcarotid/Peripheral Access Kit REF: SR-4F21G7D-MP (Kit-075-03) Sheath withAugust 19, 2026
- FDAGALT Guidewire REF SGW-051-07 Guidewires are intended for use in percutaneous procedures, to introduce andMay 14, 2025
- FDAB Braun Interventional Coaxial Dilator REF: KIT-018-47, KIT-019-67, KIT-018-41, KIT-019-37 TheseMay 14, 2025
- FDATearaway Introducer Kit, REF: KIT-010-12, KIT-010-17, KIT-900-16June 21, 2023
- FDACenteze Centesis Catheter, Catalog #: a) DRC-006-05, b) DRC-006-15,August 8, 2018
- FDAMerge Unity software, formerly known as DR Systems Unity PACS software. The firm name on the label is MergeAugust 8, 2018
Questions about this recall
Was Tearaway Introducers, Catalog #: a) BNS-000-07, b) BNS-000-10; c) DIL-015-50; d) DSS-005-11; e) DSS-005-12; f) DSS-006-045; g) DSS-007-05; h) DSS-007-165; i) DSS-010-03; j) DSS-010-04; k) DSS-010-045; l) DSS-010-05; m) DSS-012-07; n) SPT-003-46 recalled?
Yes. FDA published recall Z-2565-2018 on August 8, 2018. Reported hazard: The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter