Back to search
FDAFDA · Food and drugs·Recall number Z-2574-2026

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System

Published July 1, 2026 · Recall initiated: May 27, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Dexcom, Inc.
Distributor
Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
Lot
Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card