FDAFDA · Food and drugs·Recall number Z-2581-2026
Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAM PACK-LF DYNJ21574J JUDKINS PACK DYNJ51126
Published July 1, 2026 · Recall initiated: February 27, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
- Lot
- DYNJ21574J UDI-DI 10889942153497 Lots 21ALA586 21CLA398 21ELA408 21FLA144 21FLA788 21HLA420 21ILA431 DYNJ51126 UDI-DI 10889942122820 Lots 21ADA917
- Risk classification
- Class II