OptiLite Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages Product Usage: Used with the Odyssey 30 Holmium Laser System for fragmentation of urinary calculi and soft tissue applications including incision/excision, ablation, and coagulation
Published June 21, 2017 · Recall initiated: May 3, 2017
Summary
FDA records medical device Class II recall Z-2585-2017, published June 21, 2017 and initiated May 3, 2017. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 10,415; Lot: OptiLite Multi-Use Holmium Laser Fibers, Catalog Numbers followed by GPN: HLF-M273-CNV, G48611 HLF-M273-SMA, G48522 HLF-M365-CNV, G48613 HLF-M365-SMA, G48618 HLF-M550-CNV, G48615 HLF-M550-SMA, G48619 HLF-M940-CNV, G48617 HLF-M940-SMA, G48621 HLF-M273-CNV-CAN, G51777 HLF-M273-SMA-CAN, G51785 HLF-M365-CNV-CAN, G51778 HLF-M365-SMA-CAN, G51786 HLF-M550-CNV-CAN, G51779 HLF-M550-SMA-CAN, G51787 HLF-M940-CNV-CAN, G51780 HLF-M940-SMA-CAN, G51788. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cook Inc. is initiating a voluntary recall of the OptiLite Multi-Use Holmium Laser Fibers because it has been identified that the reprocessing instructions do not provide sufficiently detailed information for the cleaning, disinfection, and sterilization of these products.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Worldwide Distribution - US Nationwide in the states of: AK AL AZ CA CO FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC NJ NY OH OR PA SC SD TN TX WA WI Foreign: Australia, Bolivia, Brazil, Canada, Chile, Columbia, Cyprus, Czech Republic, Germany, Dominican Republic, Greece, Hong Kong, Ireland, Israel, India, Iran, Jordan, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Latvia, Libya, Mexico, Malaysia, New Zealand, Oman, Panama, Peru, Portugal, Serbia, Turkey, Trinidad and Tobago, Taiwan, and Venezuela. VA/DOD: Yes- see below
- Lot
- OptiLite Multi-Use Holmium Laser Fibers, Catalog Numbers followed by GPN: HLF-M273-CNV, G48611 HLF-M273-SMA, G48522 HLF-M365-CNV, G48613 HLF-M365-SMA, G48618 HLF-M550-CNV, G48615 HLF-M550-SMA, G48619 HLF-M940-CNV, G48617 HLF-M940-SMA, G48621 HLF-M273-CNV-CAN, G51777 HLF-M273-SMA-CAN, G51785 HLF-M365-CNV-CAN, G51778 HLF-M365-SMA-CAN, G51786 HLF-M550-CNV-CAN, G51779 HLF-M550-SMA-CAN, G51787 HLF-M940-CNV-CAN, G51780 HLF-M940-SMA-CAN, G51788
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDAPalcam Supplement Acumedia PN 7987 Product Usage: PALCAM Supplement is used with PALCAM Agar or PALCAM BrothJune 21, 2017
- FDAPointe Scientific Hitachi Total Bilirubin Rl Reagent Product Usage: Two part Chemistry Reagent Hitachi TotalJune 21, 2017
Questions about this recall
Was OptiLite Multi-Use Holmium Laser Fibers, Supplied sterilized by ethylene oxide gas in peel-open packages Product Usage: Used with the Odyssey 30 Holmium Laser System for fragmentation of urinary calculi and soft tissue applications including incision/excision, ablation, and coagulation recalled?
Yes. FDA published recall Z-2585-2017 on June 21, 2017. Reported hazard: Cook Inc. is initiating a voluntary recall of the OptiLite Multi-Use Holmium Laser Fibers because it has been identified that the reprocessing instructions do not provide sufficiently detailed information for the cleaning, disinfection, and sterilization of these products.
What should I do if I have this product?
Voluntary: Firm initiated · Letter