AXIOM Artis or Artis zee system Product Usage: Axiom Artis and Artis zee / zeego are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures
Published October 1, 2014 · Recall initiated: June 5, 2014
Summary
FDA records medical device Class II recall Z-2587-2014, published October 1, 2014 and initiated June 5, 2014. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 1067; Lot: material number 3771750, model numbers: 5904441, 5904466, 7008605, 7412807, 7413078, 7555357, 7555365, 7555373, 7727717, 7728392, 10094135, 10094137, 10094139, 10094141, 10094142, 10094143, and 10280959. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Nationwide Distribution.
- Lot
- material number 3771750, model numbers: 5904441, 5904466, 7008605, 7412807, 7413078, 7555357, 7555365, 7555373, 7727717, 7728392, 10094135, 10094137, 10094139, 10094141, 10094142, 10094143, and 10280959
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAArtis zee and Artis zeego systems. x-ray, angiographic systemOctober 1, 2014
- FDACareFusion EnVe and ReVel Ventilators. The EnVe/ ReVel Ventilator is designed for use on patients who requireOctober 1, 2014
Questions about this recall
Was AXIOM Artis or Artis zee system Product Usage: Axiom Artis and Artis zee / zeego are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures recalled?
Yes. FDA published recall Z-2587-2014 on October 1, 2014. Reported hazard: There is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.
What should I do if I have this product?
Voluntary: Firm initiated · Letter