Back to search
FDAFDA · Food and drugs·Recall number Z-2587-2014

AXIOM Artis or Artis zee system Product Usage: Axiom Artis and Artis zee / zeego are dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures

Published October 1, 2014 · Recall initiated: June 5, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential problem with the AXIOM Artis or Artis zee system in connection with the hand switch operated at the patient table which may result in uncontrolled release of an x-ray image.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution.
Lot
material number 3771750, model numbers: 5904441, 5904466, 7008605, 7412807, 7413078, 7555357, 7555365, 7555373, 7727717, 7728392, 10094135, 10094137, 10094139, 10094141, 10094142, 10094143, and 10280959
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.