Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography

Published June 21, 2017 · Recall initiated: May 8, 2017

Summary

FDA records medical device Class II recall Z-2591-2017, published June 21, 2017 and initiated May 8, 2017. Recalling firm: Merge Healthcare, Inc.; Status: Terminated; Reported quantity: 47 sites have the potentially affected versions; Lot: Versions 7.0.2.1, 7.0.2.2, 7.0.2.3, 7.0.2.3 P1, and 7.0.2.3 P2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Merge Healthcare, Inc.
Distributor
Nationwide Distribution
Lot
Versions 7.0.2.1, 7.0.2.2, 7.0.2.3, 7.0.2.3 P1, and 7.0.2.3 P2
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography recalled?

Yes. FDA published recall Z-2591-2017 on June 21, 2017. Reported hazard: When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit