BD BBL Vancomycin Screen Agar Catalog number 222204 Vancomycin Screen Agar is used to test enterococci for resistance to vancomycin and to predict the synergistic activity of this antimicrobial with an aminoglycoside antimicrobial.
Published September 24, 2014 · Recall initiated: July 28, 2014
Summary
FDA records medical device Class II recall Z-2628-2014, published September 24, 2014 and initiated July 28, 2014. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 28540; Lot: Lot numbers - 4080291 4042217 4048040 4052349 4058242 4073274 4127728. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
These lots of Agar may have exhibited breakthrough growth of vancomycin susceptible Enterococcus species control organisms. Agar that allows susceptible enterococci to grow appears as a falsely resistant culture. If a lab does not QC, the error would be identified as falsely resistant enterococci would have further AST workup. There could be a delay in the identification of false-resistance
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Worldwide Distribution: US (Nationwide) including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and WV; and Internationally to the countries of Belgium, Canada, Colombia, India, Mexico, Singapore and Taiwan.
- Lot
- Lot numbers - 4080291 4042217 4048040 4052349 4058242 4073274 4127728
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic useMay 6, 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic useMay 6, 2026
- FDAC-Series High Energy Linear Accelerator - Models: Novalis Tx, Trilogy, Trilogy Tx, Clinac iX, Clinac 2100 C/D,September 24, 2014
- FDABD Vacutainer Serum Tubes (RED TOP PLAIN TUBE 6.0 ML)Part Number: BD 367815 FMC Part Number: 87-7815-4 SingleSeptember 24, 2014
Questions about this recall
Was BD BBL Vancomycin Screen Agar Catalog number 222204 Vancomycin Screen Agar is used to test enterococci for resistance to vancomycin and to predict the synergistic activity of this antimicrobial with an aminoglycoside antimicrobial. recalled?
Yes. FDA published recall Z-2628-2014 on September 24, 2014. Reported hazard: These lots of Agar may have exhibited breakthrough growth of vancomycin susceptible Enterococcus species control organisms. Agar that allows susceptible enterococci to grow appears as a falsely resistant culture. If a lab does not QC, the error would be identified as falsely resistant enterococci would have further AST workup. There could be a delay in the identification of false-resistance
What should I do if I have this product?
Voluntary: Firm initiated · Letter