FDAFDA · Food and drugs·Recall number Z-2654-2025

Biograph mMR. Model Number: 10433372.

Published October 8, 2025 · Recall initiated: August 28, 2025

Summary

FDA records medical device Class I recall Z-2654-2025, published October 8, 2025 and initiated August 28, 2025. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Ongoing; Reported quantity: 27 units; Lot: Model Number: 10433372. UDI Number: (01)04056869006741(21)151032, (01)04056869006741(21)151057, (01)04056869006741(21)510535, (01)04056869006741(21)151034, (01)04056869006741(21)51003, (01)04056869006741(21)151021, (01)04056869006741(21)51055, (01)04056869006741(21)151013, (01)04056869006741(21)51064, (01)04056869006741(21)151044, (01)04056869006741(21)51038, (01)04056869245188(21)51006, (01)04056869006741(21)51030, (01)04056869006741(21)151004, (01)04056869006741(21)51022, (01)04056869006741(21)151049, (01)04056869006741(21)51007, (01)04056869006741(21)151045, (01)04056869006741(21)51032, (01)04056869006741(21)51057, (01)04056869006741(21)51015, (01)04056869006741(21)151025, (01)04056869006741(21)51048, (01)04056869006741(21)51062, (01)04056869006741(21)51010, (01)04056869006741(21)51051. Serial Numbers: 151032, 151057, 51053, 151034, 51003, 151021, 51055, 151013, 51064, 151044, 51038, 51006, 51030, 151004, 51022, 151049, 51007, 151045, 51032, 51057, 51015, 151025, 51048, 51062, 51010, 51051. Additional serial number added 01/23/2026: 51007. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.
Lot
Model Number: 10433372. UDI Number: (01)04056869006741(21)151032, (01)04056869006741(21)151057, (01)04056869006741(21)510535, (01)04056869006741(21)151034, (01)04056869006741(21)51003, (01)04056869006741(21)151021, (01)04056869006741(21)51055, (01)04056869006741(21)151013, (01)04056869006741(21)51064, (01)04056869006741(21)151044, (01)04056869006741(21)51038, (01)04056869245188(21)51006, (01)04056869006741(21)51030, (01)04056869006741(21)151004, (01)04056869006741(21)51022, (01)04056869006741(21)151049, (01)04056869006741(21)51007, (01)04056869006741(21)151045, (01)04056869006741(21)51032, (01)04056869006741(21)51057, (01)04056869006741(21)51015, (01)04056869006741(21)151025, (01)04056869006741(21)51048, (01)04056869006741(21)51062, (01)04056869006741(21)51010, (01)04056869006741(21)51051. Serial Numbers: 151032, 151057, 51053, 151034, 51003, 151021, 51055, 151013, 51064, 151044, 51038, 51006, 51030, 151004, 51022, 151049, 51007, 151045, 51032, 51057, 51015, 151025, 51048, 51062, 51010, 51051. Additional serial number added 01/23/2026: 51007.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Biograph mMR. Model Number: 10433372. recalled?

Yes. FDA published recall Z-2654-2025 on October 8, 2025. Reported hazard: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

What should I do if I have this product?

Voluntary: Firm initiated · Letter